Check the 3MTM RangerTM system website to ensure you have the most recent version of this document.www.rangerfluidwarming.com reorder #202524A

English

3MTM RangerTM Blood/Fluid and Irrigation Warming Systems

 

 

Model 247

Ranger Irrigation Fluid Warming, Model 247 Maintenance Log

The operating temperature and alerts should be tested at least every 6 months or according to institutional protocol.

maintenance Action Performed

model ____________ sn________

Date _ ________

Operating temperature: (Specification = 41°C±1)

______°C

 

Over-temperature alert points:

 

 

Audible

(48°C [±2] on Ranger unit display)

______°C

 

Secondary

(50°C [±2])

______°C

 

 

 

Date _ ________

Operating temperature: (Specification = 41°C±1)

______°C

 

Over-temperature alert points:

 

 

Audible

(48°C [±2] on Ranger unit display)

______°C

 

Secondary

(50°C [±2])

______°C

 

 

 

Date _ ________

Operating temperature: (Specification = 41°C±1)

______°C

 

Over-temperature alert points:

 

 

Audible

(48°C [±2] on Ranger unit display)

______°C

 

Secondary

(50°C [±2])

______°C

 

 

 

Date _ ________

Operating temperature: (Specification = 41°C±1)

______°C

 

Over-temperature alert points:

 

 

Audible

(48°C [±2] on Ranger unit display)

______°C

 

Secondary

(50°C [±2])

______°C

 

 

 

Date _ ________

Operating temperature: (Specification = 41°C±1)

______°C

 

Over-temperature alert points:

 

 

Audible

(48°C [±2] on Ranger unit display)

______°C

 

Secondary

(50°C [±2])

______°C

 

 

 

Date _ ________

Operating temperature: (Specification = 41°C±1)

______°C

 

Over-temperature alert points:

 

 

Audible

(48°C [±2] on Ranger unit display)

______°C

 

Secondary

(50°C [±2])

______°C

18