3MTM RangerTM Blood/Fluid and Irrigation Warming Systems

 

 

Specifications

physical characteristics

warming unit

4.5 in. (11 cm) high x 7.5 in. (19 cm) wide x 10 in. (25 cm) long; wt.: 7 lb. 7 oz. (3.4 kg)

certifications

IEC/EN 60601-1; CAN/CSA-C22.2, No. 601.1

classification

MEDICAL — GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1; CAN/CSA-C22.2, No. 601.1; ANSI/AAMI ES60601-1:2005 CSA-C22.2 No. 60601-1:08; Control No. 4HZ8

Classified under IEC 60601-1 Guidelines (and other national versions of the Guidelines) as Class I, Type B, Ordinary equipment, Continuous operation. Classified by Underwriters Laboratories Inc. with respect to electric shock, fire and mechanical hazards only, in accordance with IEC/EN2 60601-1 and in accordance with Canadian/CSA C22.2, No. 601.1. Classified under the Medical Device Directive as a Class IIb device.

electrical characteristics

temperature characteristics

device rating

set point temperature

100-120 VAC, 50/60 Hz41°C
220-240 VAC, 50/60 Hz

 

maximum heating power

over-temperature alert

900 WModel 245:

 

43°C

 

Model 247:

 

48°C

fuse

under-temperature alert

2 x T10A-H (250V) for 100-120 VAC33°C
2 x T6.3A-H (250V) for 220-240 VAC

 

fuse type

over-temperature cutoff

Time delay, high breaking

Model 245:

 

44°C (warming units Rev N and newer)

 

46°C (warming units Rev A to M)

 

Model 247:

 

50°C

leakage current

Meets leakage current requirements in accordance with UL/IEC 60601-1.

environmental conditions

operating temperature range

15° to 40°C (59° to 104°F)

storage temperature range

-20° to 45°C (-4° to 113°F)

operating humidity

10 to 85% RH, noncondensing

atmospheric pressure range

50 kPa to 106 kPa

22