Replacing the Power Fuse
The power fuses are located in the power entry module.tools and equipment
Small slotted screwdrivermethod
1.Disconnect the warming unit from the power source.2.Remove the power cord.3.Locate the fuse carrier in the power entry module.4.Use the small screwdriver to remove the fuse carrier from the power entry module.
5.Remove the blown fuse from the fuse carrier and replace with a new fuse (as marked).
6.Place the fuse carrier back into the power entry module.7.Reattach the power cord.8.Reconnect the warming unit to the power source.9.Turn the unit on to verify normal operation.10.Record the maintenance action taken.19
202524A