3MTM RangerTM Blood/Fluid and Irrigation Warming Systems

 

 

Introduction

The Ranger warming system includes a warming unit and a disposable warming set. The Ranger warming unit is designed to be mounted to an I.V. pole. A handle located on the top of the unit makes transport easy. When mounted to the I.V. pole, the unit fits easily above the 3MTM Bair HuggerTM 500 or 700 series warming unit. For additional information on Ranger warming sets visit us online at rangerfluidwarming.com.

Model 245 Blood/Fluid Warming System

The Ranger blood/fluid warming system is designed to warm blood, blood products, and liquids and deliver these at flow rates from KVO to 500 mL/min. At these flow rates, the device maintains fluid output temperatures ranging from 33°C to 41°C (Note: This is for room temperature fluids only). It takes less than two minutes to warm up to the 41°C ±1 set point temperature. The alert points on the Model 245 are 43°C ±1 and 46°C [+3, -2°C].

Disposable warming sets are available for standard flow, high flow, and pediatric applications. Warming sets are sterile, latex-free, single-use-only items and are designed to be used with the reusable warming unit. Do not use other warming sets with the Model 245, including those designed for the Model 247 irrigation fluid warming system.

Model 247 Irrigation Fluid Warming System

The Ranger irrigation fluid warming system is designed to warm irrigation fluids and deliver these at flow rates from 0 to 865 mL/min when the bag is hung 100 cm above the scope. At these flow rates, the device maintains fluid output temperatures ranging from 33°C to 41°C (Note: This is for room temperature fluids only). It takes less than two minutes to warm up to the 41°C ±1 set point temperature. The alert points on the Model 247 are 48°C (±2) and 50°C (±2).

The Ranger irrigation disposable fluid warming set is sterile, latex-free, and is designed to be used with the Ranger irrigation fluid warming unit. Do not use other warming sets with the Model 247, including those designed for the Model 245 blood/fluid warming unit.

Indications for Use

Blood/Fluid Warming System

The Ranger blood/fluid warming system is intended to warm blood, blood products, and liquids.

Irrigation Fluid Warming System

The Ranger irrigation fluid warming system is intended to warm irrigation fluids.

Definition of Symbols

ON

OFFFuseCAUTIONType B Applied Part