Check the 3MTM RangerTM system website to ensure you have the most recent version of this document.www.rangerfluidwarming.com reorder #202524A

English

3MTM RangerTM Blood/Fluid and Irrigation Warming Systems

 

 

Ranger Irrigation Fluid Warming System, Model 247 Preventative

Maintenance Checklist

Caution

Temperature adjustments cannot be made to the Ranger irrigation fluid warming unit. If the test shows the temperature is out of specification, please call 3M Patient Warming technical support at 1-800-733-7775.

Tools and equipment

Inlet tube

Thermocouple: 0.005" lead or smaller(part #90020)

Model 247

Calibrated thermocouple temperature monitor

Aluminum tape

Ranger irrigation fluid warming set

Phillips screwdriver

Saline/water

1.

Use aluminum tape to affix a thermocouple

 

to the top surface of the Ranger irrigation

 

fluid warming cassette as shown in Fig. 1.

2.

Slide the warming cassette into the slot of

 

the Ranger irrigation fluid warming unit.

3.

Using room temperature saline/

 

water, prime the cassette per

 

instructions included with it.

 

note: To achieve correct results,

 

fluids must be administered into the

2mL Draincassette Insert cassette from .priming before 247 l de Mo

Warming Fluid Irrigation

more or remove to

 

.

 

Aluminum tape

Thermocouple

0.75"

8.50"

Fig. 1: Ranger irrigation fluid warming cassette

Fig. 2: Ranger irrigation fluid warming set

 

inlet tube, NOT into the outlet tube.

4.

Close patient line with white clamp

 

to stop flow.

5.

Turn the Ranger irrigation

 

warming unit ON and allow it to

 

warm up for 5 minutes.

6.

The Ranger irrigation warming

 

unit display should read 41°C±1.

 

Note the temperature on the

 

thermocouple temperature

Fluid source

Inlet tube

Cassette

White clamp

monitor. If the temperature reading
is not 41°C±1, call 3M Patient
Warming technical support.
note: Most handheld temperature
meters have a ±1°C tolerance.

Outlet tube

Blue clamp

Exit

16