Check the 3MTM RangerTM system website to ensure you have the most recent version of this document.www.rangerfluidwarming.com reorder #202524A

3MTM RangerTM Blood/Fluid and Irrigation Warming Systems

English

 

 

Ranger Blood/Fluid Warming Unit, Model 245 Maintenance Log

The operating temperature and alerts should be tested at least every 6 months or according to institutional protocol.

maintenance Action Performed

model ____________ sn_________

Date _ _________

Operating temperature: (Specification = 41°C±1)

______°C

 

Over-temperature alert points:

 

 

Audible

(43°C [±1] on Ranger unit display)

______°C

 

Secondary

(46°C [+3, -2])

______°C

 

 

 

Date _ _________

Operating temperature: (Specification = 41°C±1)

______°C

 

Over-temperature alert points:

 

 

Audible

(43°C [±1] on Ranger unit display)

______°C

 

Secondary

(46°C [+3, -2])

______°C

 

 

 

Date _ _________

Operating temperature: (Specification = 41°C±1)

______°C

 

Over-temperature alert points:

 

 

Audible

(43°C [±1] on Ranger unit display)

______°C

 

Secondary

(46°C [+3, -2])

______°C

 

 

 

Date _ _________

Operating temperature: (Specification = 41°C±1)

______°C

 

Over-temperature alert points:

 

 

Audible

(43°C [±1] on Ranger unit display)

______°C

 

Secondary

(46°C [+3, -2])

______°C

 

 

 

Date _ _________

Operating temperature: (Specification = 41°C±1)

______°C

 

Over-temperature alert points:

 

 

Audible

(43°C [±1] on Ranger unit display)

______°C

 

Secondary

(46°C [+3, -2])

______°C

 

 

 

Date _ _________

Operating temperature: (Specification = 41°C±1)

______°C

 

Over-temperature alert points:

 

 

Audible

(43°C [±1] on Ranger unit display)

______°C

 

Secondary

(46°C [+3, -2])

______°C

245 Model

15

202524A