6-2. Function list

Item

Function number

 

Description

 

 

Š C01- 0

CAL switch enabled,

0

key enabled

Do not set

Calibration

C01- 1

CAL switch disabled,

0

key disabled

“C01-1”

 

C01- 2

CAL switch enabled,

0

key disabled

 

 

 

Š C02- 0

6 kg (15 lb)

 

 

 

 

C02- 1

15 kg (35 lb)

 

 

 

 

C02- 2

30 kg (70 lb)

 

 

 

Capacity

C02- 3

60 kg (150 lb)

 

 

 

C02- 4

150 kg (350 lb)

 

 

 

 

 

 

 

 

C02- 5

300 kg (700 lb)

 

 

 

 

C02- 6

600 kg (1500 lb)

 

 

 

 

C02- 7

1500 kg (3500 lb)

 

 

 

Weighing display

C03- 0

kg, g

 

 

See 11-1.

unit

Š C03- 1

kg, g, lb, oz, lb-oz

 

 

Capacity ...”.

Power-on-zero

C04- 0

±10% of the capacity

 

 

OIML R76 / NTEP

range

Š C04- 1

±50% of the capacity

 

 

 

Internal use

Š C05- 0

Do not change.

 

 

 

Decimal

Š C06- 0

. (point)

 

 

 

separator

C06- 1

, (comma)

 

 

 

Tare clear by

Š C07- 0

ZERO does not clear tare weight

 

ZERO operation

C07- 1

ZERO clears tare weight

 

OIML R76

Tare operation

C08- 0

Tare weight is rounded to the weight

NTEP

(Net weight

display count

 

 

 

 

 

 

calculation)

Š C08- 1

Internal count is used for TARE weight.

OIML R76

Preset tare

C09- 0

Function disabled

 

 

 

function

Š C09- 1

Function enabled

 

 

 

lb-oz unit

C10- 0

lb-oz disabled

 

 

 

Š C10- 1

lb-oz enabled

 

 

 

 

 

 

 

ZERO operation

Š C11- 0

±2% of the capacity

 

 

 

range

C11- 1

Up to the capacity

 

 

 

 

Š C12- 0

“f02” setting specifies the resolution.

 

Display resolution

C12- 1

Fixed to “f02-0” (1/3,000~1/3,500)

 

C12- 2

Fixed to “f02-1” (1/6,000~1/7,500)

 

 

 

 

C12- 3

Fixed to “f02-2” (1/12,000~1/15,000)

 

Auto-tare function

C13- 0

Function disabled

 

 

C13-0 deletes

Š C13- 1

Function enabled

 

 

“f21” ~ “f23”.

ŠFactory setting

11

Page 13
Image 13
A&D 1WMPD4002313 manual Function list