10. ERROR MESSAGES

Overload error

----e---

Power-on error

--------

Low battery

---lb1--

CAL error

---Cal-e

Other errors

---err-#

Warning to indicate that an object beyond the weighing capacity has been placed on the weighing pan. Remove the object from the weighing pan.

Warning to indicate that an object beyond the power-on zero range (50% of the weighing capacity or 10%) is placed on the weighing platform. Remove the object from the weighing pan.

Pressing the PRINT/ENT key, the indicator will show the weight value without zeroing the indicator.

Warning to show that the OP-02 battery is discharged. Immediately connect the power cord to the AC power source to charge the battery. The indicator can be used while the battery is charging (except when the OP-04 is installed).

Warning to indicate that the calibration has been canceled because the calibration weight is too light. Check if the calibration weight is correct.

Showing errors inside the indicator. Check the connection between the indicator and platform, and turn the power off and on with nothing on the weighing platform. (# signifies an error number.)

When the indicator does not recover from errors or shows other errors, contact your local A&D dealer.

18

Page 20
Image 20
A&D 1WMPD4002313 manual Error Messages, Lb1, Err-#