5725A Instruction Manual

 

 

 

Table 7-1. 5725A Fault Codes

 

 

 

 

 

 

FAULT

 

 

MESSAGE

 

EXPLANATION

NO.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

201

5725

Self-Test ROM Failure

 

Requires service

202

5725

Self-Test RAM Failure

 

Requires service

 

203

5725

Self-Test EEPROM Failure

 

Requires service

 

204

5725

Self-Test Data Bus Failure

 

Requires service

 

205

5725

Self-Test CLAMPS Circuit Failure

 

Requires service

 

206

5725

Self-Test HVCLR Circuit Failure

 

Requires service

 

207

5725

Self-Test DAC Failure

 

Requires service

 

208

5725

Self-Test Watchdog Timer Failure

 

Requires service

 

209

5725

Current Heatsink Too Hot

 

Requires service

 

210

 

Output Tripped To Standby

 

Current heat sink too hot in operate

 

 

211

5725

Current Compliance Voltage Too

 

Internal fault, requires service

 

 

 

High

 

 

 

212

5725

Current Compliance Voltage Too

 

Most likely an external fault

 

 

 

High

 

 

 

213

5725

+400V Supply Did Not Shut Off

 

Requires service

 

214

5725

-400V Supply Did Not Shut Off

 

Requires service

 

215

5725

Voltage Heatsink Too Hot

 

Requires service

 

216

5725

Voltage Heatsink Too Hot

 

Voltage heat sink too hot in operate

 

217

5725

+400V Supply Too Small

 

Requires service, or improper line voltage

 

 

 

 

 

 

selected

218

5725

+400V Supply Too Large

 

Requires service, or improper line voltage

 

 

 

 

 

 

selected

219

5725

-400V Supply Too Large

 

Requires service, or improper line voltage

 

 

 

 

 

 

selected

220

5725

-400V Supply Too Small

 

Requires service, or improper line voltage

 

 

 

 

 

 

selected

221

5725

+400V Supply Current Too High

 

Internal fault, requires service

 

222

 

Output Tripped To Standby

 

+400V supply too high in operate, most likely an

 

 

 

 

 

 

 

external fault

223

5725

-400V Supply Current Too High

 

Internal fault, requires service

 

224

 

Output Tripped To Standby

 

-400V supply magnitude too high in operate,

 

 

 

 

 

 

 

most likely an external fault

225

5725

Fan Not Working

 

Requires service

 

226

5725

CLAMPS Fault

 

Internal fault, requires service

 

227

 

Output Tripped To Standby

 

CLAMPS circuit activated most likely because of

 

 

 

 

 

 

 

external transient condition

 

 

 

 

 

 

7-4

Page 120
Image 120
Fluke 5725A instruction manual A Fault Codes, Fault Message Explanation

5725A specifications

The Fluke 5725A is a high-performance, multifunctional temperature calibration source designed to meet the demanding requirements of laboratory and industrial environments. Renowned for its accuracy, reliability, and versatility, the 5725A serves as a powerful tool for engineers and technicians engaged in temperature calibration processes.

At the core of the 5725A is its advanced measurement technology, which allows for precision temperature calibrations across a wide range. With an operating temperature range from -200°C to 660°C, it caters to various applications, making it suitable for calibrating thermocouples, resistance temperature detectors (RTDs), and other temperature measurement devices.

One of the standout features of the Fluke 5725A is its outstanding accuracy, boasting a specified uncertainty of just ±0.15°C. This level of precision ensures that users can maintain compliance with regulatory standards and regulatory requirements in fields such as aerospace, pharmaceuticals, and manufacturing.

The Fluke 5725A is equipped with dual-channel capabilities, allowing users to calibrate two devices simultaneously. This enhances productivity by minimizing downtime and optimizing workflow efficiency. The intuitive touchscreen interface simplifies operations, enabling users to view measurements, set up test parameters, and access historical data with ease.

Versatile connectivity options such as USB and Ethernet enable seamless integration into existing systems and allow for data transfer, remote control, and configuration. This feature empowers technicians to conduct calibrations efficiently, whether on-site or remotely.

The robust design of the Fluke 5725A ensures its durability in demanding environments. Built with high-quality materials, it is resistant to shocks and vibrations, making it suitable for use in manufacturing plants and laboratories where equipment may be subjected to harsh conditions.

Another notable characteristic of the 5725A is its ability to generate stable temperature outputs with minimal drift, contributing to consistency in calibration results. Its built-in self-test functionality helps ensure that the unit remains in peak operating condition, giving users confidence in their calibration processes.

In summary, the Fluke 5725A is a top-tier temperature calibration source that excels in accuracy, reliability, and user-friendliness. Whether in laboratory settings or industrial applications, its advanced technologies and features make it an essential tool for professionals tasked with maintaining precision in temperature measurements.