Manual Supplement

Change #1

On page 6-8, replace the first sentence in the Warning with the last sentence in step 6.

On page 6-9, paragraph 6-8, replace step 2 with:

2.Return the High Voltage Amplifier assembly (A3) (and the Current Amplifier assembly (A2) if it has been accessed) to the normal position, and reinstall the assembly retaining screws.

Replace the last Note with:

Note

When reinstalling the assembly, lower it into place squarely to make sure its connector seats correctly in the Interconnect assembly connector. Ensure that both J301 and J303 are properly aligned and connected to their respective cables.

On page 6-11, paragraph 6-10, following the WARNING statement:

Change: The Current Amplifier assembly lays flat on the top of the instrument, toward the

front...

To: The Current Amplifier assembly lays flat on the top of the instrument, toward the

rear...

On page 6-12, paragraph 6-11, change the first sentence:

From: The High Voltage Amplifier assembly lays flat on the top of the instrument, toward the rear.

To: The High Voltage Amplifier assembly lays flat on the top of the instrument, toward the front.

On page 6-13, paragraph 6-12, make the following changes:

Change: The High Voltage Sense assembly lays flat on the top of the instrument

To: The High Voltage Sense assembly lays flat on the bottom of the instrument Change step 2 to step 2a and add a new step 2 as follows:

2.Follow the Current Amplifier Assembly (A2) access procedure to gain access to the connectors on the High Voltage Sense Assembly (A6). Remove the connectors and secure the Current Amplifier Assembly before proceeding.