5725A
Instruction Manual
Accessing the Fuse |
wCaution
To prevent instrument damage, verify that the correct fuse is installed for the line voltage setting. Acceptable fuse types are Bussman type MTH or Littelfuse type 312 or equivalent, of the appropriate amperage rating. Use only a 4A, 250V
Note
A spare 4A, 250V fuse is shipped with the 5725A to ensure that one is available for use on the
The line power fuse is accessible on the rear panel. The fuse rating label to the right of the fuse holder (labeled FUSE) shows the correct replacement fuse rating for each line voltage setting. To check or replace the fuse, refer to Figure
1.Turn off the POWER switch and disconnect the line power cord from ac line power.
2.Insert the blade of a standard screwdriver in the slot of the fuse holder labeled FUSE.
3.Turn the screwdriver counterclockwise until the cap and fuse pop free.
4.To reinstall the fuse, reverse this procedure.
WARNING GROUNDING
CONNECTOR IN POWER CORD
MUST BE CONNECTED TO
ENSURE PROTECTION FROM
ELECTRONIC SHOCK.
!
FUSE
CHASSIS
GROUND
S2
CAUTION FOR FIRE PROTECTION REPLACE ONLY WITH A 250V FUSE OF INDICATED RATING.
VOLTAGE | FUSE | |
SELECTION | ||
|
S2 | S3 | S4 |
100V |
| ! |
110V |
| F8A 250V |
115V |
| (FAST) |
120V |
|
|
S2 | S3 | S4 |
200V |
| ! |
220V |
| F4A 250V |
230V |
| (FAST) |
240V |
|
|
S3 |
| S4 |
JOHN FLUKE MFG. CO., INC.
MADE IN U.S.A.
PATENTS PENDING
NO INTERNAL USER SERVICEABLE | CURRENT | |||||||
OUTPUT | ||||||||
PARTS REFER SERVICE TO | ||||||||
QUALIFIED SERVICE PERSONNEL |
| |||||||
TO CLEAN FILTER |
| |||||||
REMOVE FROM INSTRUMENT | HI | |||||||
AND FLUSH WITH WARM |
| |||||||
SOAPY WATER | 11A MAX | |||||||
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
| |
|
|
|
|
| 20V PK | LO | ||
|
|
|
|
|
| MAX |
| |
|
|
|
|
|
|
|
| |
|
|
|
|
|
| |||
|
|
|
| |||||
5700A CALIBRATOR |
|
CALIBRATION
ENABLE NORMAL
aq5f.eps
Figure