Introduction

Introduction

5-1.

The 5725A is calibrated at the factory before shipping. Calibration is traceable to the U.S. National Bureau of Standards. All that is required to maintain traceability is to run 5700A calibration to external standards at the beginning of the calibration cycle and do performance verification every two years. Calibration check and range calibration are optional procedures that are available for special needs. The 5725A is calibrated whenever an attached 5700A is calibrated.

Where to Find Further Information

5-2.

Information about 5700A calibration, and therefore 5725A calibration, is contained in the following sections of the 5700A manual set:

Section 7 of the 5700A Operator Manual presents procedures for calibration, calibration check, and range calibration.

Section 1 of the 5700A Operator Manual describes the calibration process and the theory behind its use to establish traceability to national standards. The same section includes a description of the calibration check feature, and suggests using it to develop a performance history for your 5700A. (This applies equally to a 5725A.)

Section 3 of the 5700A Service Manual contains a performance verification procedure recommended every two years to maintain traceability.