Troubleshooting 7
Testing the Power Supply By Itself
3.Remove plug P301 from J301 and inserting it into J303. Now the high voltage power supply will power up in the ±50V mode. The high voltage amplifier may also be serviced in the ±50V mode.
Testing the Power Supply By Itself |
|
To turn on a Power Supply assembly (A4) by itself for troubleshooting, proceed as follows:
1. Turn off the POWER switch and disconnect the line power cord from ac line power.
Warning
Follow the instructions under "initial access procedure" in section 6 to make sure high voltage has been bled off before you touch anything inside the 5725A.
The power supply assembly generates lethal voltages. Use extreme caution when you are working inside the 5725A with the power connected. Use only
2.Place the Power Supply assembly in the service position as described in the access procedures in Section 6.
3.Plug P301 onto the E302 header where the label MAN appears (pins 2 and 3).
4.This allows
5.You may want to troubleshoot in the