5725A

Instruction Manual

AC Voltage

Range Resolution

Frequency

Hz

Absolute Uncertainty

±5°C from calibration temperature

24 Hours

90 Days

180 Days

1 Year

±(ppm output + μV)

Relative Uncertainty

±1°C

24 Hours

90 Days

±(ppm output + μV)

 

2.2 mV

 

1 nV

 

 

 

 

 

 

 

 

 

22 mV

10 nV

 

 

220 mV

100 nV

 

 

2.2V

1 μV

 

 

22V

10 μV

 

 

10-20

20-40

40-20k

20k-50k

50k-100k

100k-300k

300k-500k

500k-1M

10-20

20-40

40-20k

20k-50k

50k-100k

100k-300k

300k-500k

500k-1M

10-20

20-40

40-20k

20k-50k

50k-100k

100k-300k

300k-500k

500k-1M

10-20

20-40

40-20k

20k-50k

50k-100k

100k-300k

300k-500k

500k-1M

10-20

20-40

40-20k

20k-50k

50k-100k

100k-300k

300k-500k

500k-1M

500 + 5

200 + 5

100 + 5

340 + 5

800 + 8

.11% + 15

.15% + 30

.4% + 40

500 + 6

200 + 6

100 + 6

340 + 6

800 + 8

.11% + 15

.15% + 30

.4% + 40

500 + 16

200 + 10

95 + 10

300 + 10

750 + 30

940 + 30

.15% + 40

.30% + 100

500 + 100

150 + 30

70 + 7

120 + 20

230 + 80

400 + 150

.10% + 400

.20% + 1 mV

500 + 1 mV

150 + 300

70 + 70

120 + 200

230 + 400

500 + 1.7 mV

.12% + 5 mV

.26% + 9 mV

550 + 5

220 + 5

110 + 5

370 + 5

900 + 8

.12% + 15

.17% + 30

.44% + 4

550 + 6

220 + 6

110 + 6

370 + 6

900 + 8

.12% + 15

.17% + 30

.44% + 4

550 + 16

220 + 10

100 + 10

330 + 10

800 + 30

.1% + 30

.17% + 40

.33% + 100

550 + 100

170 + 30

75 + 7

130 + 20

250 + 80

440 + 150

.11% + 400

.22% + 1 mV

550 + 1 mV

170 + 300

75 + 70

130 + 200

250 + 400

550 + 1.7 mV

.13% + 5 mV

.28% + 9 mV

600 + 5

230 + 5

120 + 5

390 + 5

950 + 8

.13% + 15

.17% + 30

.47% + 40

600 + 6

230 + 6

120 + 6

390 + 6

950 + 8

.13% + 15

.17% + 30

.47% + 40

600 + 16

230 + 10

110 + 10

350 + 10

850 + 30

.11% + 30

.17% + 40

.35% + 100

600 + 100

170 + 30

80 + 7

140 + 20

270 + 80

470 + 150

.12% + 400

.23% + 1 mV

600 + 1 mV

170 + 300

80 + 70

140 + 200

270 + 400

550 + 1.7 mV

.13% + 5 mV

.29% + 9 mV

600 + 5

240 + 5

120 + 5

410 + 5

950 + 8

.13% + 15

.18% + 30

.48% + 40

600 + 6

240 + 6

120 + 6

410 + 6

950 + 8

.13% + 15

.18% + 30

.48% + 40

600 + 16

240 + 10

110 + 10

360 + 10

900 + 30

.11% + 30

.18% + 40

.36% + 100

600 + 100

180 + 30

85 + 7

140 + 20

280 + 80

480 + 150

.12% + 400

.24% + 1 mV

600 + 1 mV

180 + 300

85 + 70

140 + 200

280 + 400

600 + 1.7 mV

.14% + 5 mV

.30% + 9 mV

500 + 5

200 + 5

60 + 5

100 + 5

220 + 8

400 + 15

.10% + 30

.3% + 30

500 + 6

200 + 6

60 + 6

100 + 6

220 + 8

400 + 15

.10% + 30

.3% + 30

500 + 16

200 + 10

60 + 10

100 + 10

220 + 30

400 + 30

.1% + 40

.3% + 100

500 + 100

150 + 30

40 + 7

100 + 20

200 + 80

400 + 150

.1% + 400

.2% + 1 mV

500 + 1 mV

150 + 300

40 + 70

100 + 200

200 + 400

500 + 1.7 mV

.12% + 5 mV

.26% + 9 mV

550 + 5

220 + 5

65 + 5

110 + 5

240 + 8

440 + 15

.11% + 30

.33% + 30

550 + 6

220 + 6

65 + 6

110 + 6

240 + 8

440 + 15

.11% + 30

.33% + 30

550 + 16

220 + 10

65 + 10

110 + 10

240 + 30

440 + 30

.11% + 40

.33% + 100

550 + 100

170 + 30

45 + 7

110 + 20

220 + 80

440 + 150

.11% + 400

.22% + 1 mV

550 + 1 mV

170 + 300

45 + 70

110 + 200

220 + 400

550 + 1.7 mV

.13% + 5 mV

.28% + 9 mV

±(ppm output + mV)

±(ppm output + mV)

220V

100 μV

 

 

1100V

11 mV

5725A Amplifier:

1100V

1 mV

750V

10-20

20-40

40-20k

20k-50k

50k-100k

100k-300k

300k-500k

500k-1M

50-1k

40-1k

1k-20k

20k-30k

30k-50k

50k-100k

500 + 10

150 + 3

75 + 1

200 + 4

500 + 10

.15% + 110

.50% + 110

1.20% + 220

75 + 4

75 + 4

105 + 6

230 + 11

230 + 11

600 + 45

550 + 10

170 + 3

80 + 1

220 + 4

550 + 10

.15% + 110

.52% + 110

1.25% + 220

80 + 4

80 + 4

125 + 6

360 + 11

360 + 11

.13% + 45

600 + 10

170 + 3

85 + 1

240 + 4

600 + 10

.16% + 110

.53% + 110

1.25% + 220

85 + 4

85 + 4

135 + 6

440 + 11

440 + 11

.16% + 45

600 + 10

180 + 3

90 + 1

250 + 4

600 + 10

.16% + 110

.54% + 110

1.30% + 220

90 + 4

90 + 4

165 + 6

600 + 11

600 + 11

.23% + 45

500 + 10

150 + 3

45 + 1

100 + 1

300 + 10

.15% + 110

.50% + 110

1.20% + 220

50 + 4

50 + 4

85 + 6

160 + 11

160 + 11

380 + 45

550 + 10

170 + 3

50 + 1

110 + 1

330 + 10

.15% + 100

.52% + 110

1.20% + 220

55 + 4

55 + 4

105 + 6

320 + 11

320 + 11

.12% + 45

1-8

Page 24
Image 24
Fluke instruction manual AC Voltage, 5725A Amplifier

5725A specifications

The Fluke 5725A is a high-performance, multifunctional temperature calibration source designed to meet the demanding requirements of laboratory and industrial environments. Renowned for its accuracy, reliability, and versatility, the 5725A serves as a powerful tool for engineers and technicians engaged in temperature calibration processes.

At the core of the 5725A is its advanced measurement technology, which allows for precision temperature calibrations across a wide range. With an operating temperature range from -200°C to 660°C, it caters to various applications, making it suitable for calibrating thermocouples, resistance temperature detectors (RTDs), and other temperature measurement devices.

One of the standout features of the Fluke 5725A is its outstanding accuracy, boasting a specified uncertainty of just ±0.15°C. This level of precision ensures that users can maintain compliance with regulatory standards and regulatory requirements in fields such as aerospace, pharmaceuticals, and manufacturing.

The Fluke 5725A is equipped with dual-channel capabilities, allowing users to calibrate two devices simultaneously. This enhances productivity by minimizing downtime and optimizing workflow efficiency. The intuitive touchscreen interface simplifies operations, enabling users to view measurements, set up test parameters, and access historical data with ease.

Versatile connectivity options such as USB and Ethernet enable seamless integration into existing systems and allow for data transfer, remote control, and configuration. This feature empowers technicians to conduct calibrations efficiently, whether on-site or remotely.

The robust design of the Fluke 5725A ensures its durability in demanding environments. Built with high-quality materials, it is resistant to shocks and vibrations, making it suitable for use in manufacturing plants and laboratories where equipment may be subjected to harsh conditions.

Another notable characteristic of the 5725A is its ability to generate stable temperature outputs with minimal drift, contributing to consistency in calibration results. Its built-in self-test functionality helps ensure that the unit remains in peak operating condition, giving users confidence in their calibration processes.

In summary, the Fluke 5725A is a top-tier temperature calibration source that excels in accuracy, reliability, and user-friendliness. Whether in laboratory settings or industrial applications, its advanced technologies and features make it an essential tool for professionals tasked with maintaining precision in temperature measurements.