Fluke 5725A Introduction and Specifications, DC Voltage Range Resolution, Absolute Uncertainty

Models: 5725A

1 196
Download 196 pages 44.16 Kb
Page 23
Image 23
Specifications

Introduction and Specifications 1

Specifications

DC Voltage

Range Resolution

Absolute Uncertainty

±5°C from calibration temperature

24 Hours

90 Days

180 Days

1 Year

 

 

 

 

±(ppm output + μV)

Relative Uncertainty

±1°C

24 Hours

90 Days

 

 

±(ppm output + μV)

220 mV

10 nV

6.5 + .75

7 + .75

8 + .75

9 + .8

2.2V

100 nV

3.5 + 1.2

6 + 1.2

7 + 1.2

8 + 1.2

11V

1

μV

3.5 + 3

5 + 4

7 + 4

8 + 4

22V

1

μV

3.5 + 6

5 + 8

7 + 8

8 + 8

220V

10

μV

5 + 100

6 + 100

8 + 100

9 + 100

1100V

100

μV

7 + 600

8 + 600

10 + 600

11 + 600

2.5 + .5

4 + .5

2.5 + 1.2

4 + 1.2

1.5 + 3

3.5 + 4

1.5 + 6

3.5 + 8

2.5 + 100

4 + 100

3 + 600

4.5 + 600

Secondary Performance Specifications and Operating Characteristics

Included in Uncertainty Specifications

Stability

Temperature Coefficient

[Note]

Noise

Range

220mV 2.2V 11V 22V

220V

1100V

±1°C

24 Hours

±(ppm output + μV)

.3 + .3

.3 + 1

.3 + 2.5

.4 + 5

.5 + 40

.5 + 200

 

0°-10°C

10°-40°C

and

 

 

40°-50°C

 

 

±(ppm output + μV)/°C

.4 + .1

1.5 + .5

.3 + .1

1.5 + 2

.15 + .2

1 + 1.5

.2 + .4

1.5 + 3

.3 + 5

1.5 + 40

.5 + 10

3 + 200

Linearity

Bandwidth

± 1°C

0.1-10 Hz

 

 

 

 

pk-pk

 

 

±(ppm output + μV)

1 + .2

.15 + .1

1 + .6

.15 + .4

.3 + 2

.15 + 2

.3 + 4

.15 + 4

1 + 40

.15 + 60

1 + 200

.15 + 300

Bandwidth 10-10 kHz

RMS

μV

5

15

50

50

150

500

Note: Temperature coefficient is an adder to uncertainty specifications that does not apply unless operating more than ±5°C from calibration

temperature.

Minimum output: 0V for all ranges, except 100V for 1100V range

Maximum load: 50 mA for 2.2V through 220V ranges; 20 mA for 1100V range; 50Ω output impedance on 220 mV range; all ranges <1000 pF,>25Ω

Load regulation: <0.2 ppm + 0.2 μV change, full load to no load

Line regulation: <0.1 ppm change, ±10% of selected nominal line

Settling time: 3 seconds to full accuracy; + 1 second for range or polarity change; + 1 second for 1100V range

Overshoot: <5%

Common mode rejection: 140 dB, DC to 400 Hz

Remote sensing: Available 0V to ±1100V, on 2.2V through 1100V ranges

1-7

Page 23
Image 23
Fluke 5725A Introduction and Specifications, DC Voltage Range Resolution, Absolute Uncertainty, Hours, Days, Year, Noise