5725A

Instruction Manual

Wideband AC Voltage (Option -03)

Specifications apply to the end of the cable and 50Ω termination used for calibration:

Range

 

 

Absolute Uncertainty

 

±5°C from calibration temperature

 

 

 

30 Hz-500 kHz

Volts

dBm

Resolution

24 Hours

90 Days

180 Days

±(% output + μV)

1 Year

1.1mV 3 mV

11mV

33mV

110mV

330mV

1.1V

3.5V

-46 -37 -26 -17

-6.2

+3.4

+14

+24

10nV

10nV

100 nV

100nV

1 μV

1 μV

10μV

10μV

.4 + .4

.4 + 1

.2 + 4

.2 + 10

.2 + 40

0.2+ 100

.2 + 400

15 + 500

.5 + .4

.45 + 1

.35 + 4

.3 + 10

.3 + 40

0.25+ 100

.25 + 400

.2 + 500

.6 + .4

.5 + 1

.5 + 4

.45 + 10

.45 + 40

.35 + 100

.35 + 400

.3 + 500

.8 + 2

.7 + 3

.7 + 8

.6 + 16

.6 + 40

.5 + 100

.5 + 400

.4 + 500

 

 

Amplitude Flatness, 1 kHz

 

 

Frequency

Reference

Temperature

Frequency

 

Resolution

Voltage Range

Coefficient

 

 

 

 

 

 

 

Settling

Time To

Full

Accuracy

Harmonic Distortion

Hz

10-30

30-120

120-1.2k

1.2k-12k

12k-120k

120k-1.2M

1.2M-2M

2M-10M

10M-20M

20M-30M

Hz

.01

.01

.1

1

10

100

100k

100k

1M

1M

1.1 mV

.3

.1

.1

.1

.1

.2 + 3 μV

.2 + 3 μV

.4 + 3 μV

.6 + 3 μV

1.5 + 15 μV

3mV

±%

.3

.1

.1

.1

.1

.1 + 3 μV

.1 + 3 μV

.3 + 3 μV

.5 + 3 μV

1.5 + 3 μV

>3 mV

.3

.1

.1

.1

.1

.1 + 3 μV

.1 + 3 μV

.2 + 3 μV

.4 + 3 μV

1 + 3 μV

±ppm/°C

100

100

100

100

100

100

100

100

150

300

Seconds

7

7

5

5

5

5

0.5

0.5

0.5

0.5

dB

-40

-40

-40

-40

-40

-40

-40

-40

-34

-34

Additional Operating Information:

dBm reference = 50Ω

Range boundaries are at voltage points, dBm levels are approximate.

dBm = 10 log ( Power ),

1 mW

0.22361V across 50Ω = 1 mW or 0 dBm

Minimum output: 300 μV (-57 dBm)

Frequency uncertainty: ±0.01%

Frequency resolution: 11,999 counts to 1.1999 MHz, 119 counts to 30 MHz.

Overload protection: A short circuit on the wideband output will not result in damage. After settling time, normal operation is restored upon removal.

1-16

Page 32
Image 32
Fluke 5725A Wideband AC Voltage Option, Range Absolute Uncertainty, Hz-500 kHz, Amplitude Flatness, 1 kHz Frequency

5725A specifications

The Fluke 5725A is a high-performance, multifunctional temperature calibration source designed to meet the demanding requirements of laboratory and industrial environments. Renowned for its accuracy, reliability, and versatility, the 5725A serves as a powerful tool for engineers and technicians engaged in temperature calibration processes.

At the core of the 5725A is its advanced measurement technology, which allows for precision temperature calibrations across a wide range. With an operating temperature range from -200°C to 660°C, it caters to various applications, making it suitable for calibrating thermocouples, resistance temperature detectors (RTDs), and other temperature measurement devices.

One of the standout features of the Fluke 5725A is its outstanding accuracy, boasting a specified uncertainty of just ±0.15°C. This level of precision ensures that users can maintain compliance with regulatory standards and regulatory requirements in fields such as aerospace, pharmaceuticals, and manufacturing.

The Fluke 5725A is equipped with dual-channel capabilities, allowing users to calibrate two devices simultaneously. This enhances productivity by minimizing downtime and optimizing workflow efficiency. The intuitive touchscreen interface simplifies operations, enabling users to view measurements, set up test parameters, and access historical data with ease.

Versatile connectivity options such as USB and Ethernet enable seamless integration into existing systems and allow for data transfer, remote control, and configuration. This feature empowers technicians to conduct calibrations efficiently, whether on-site or remotely.

The robust design of the Fluke 5725A ensures its durability in demanding environments. Built with high-quality materials, it is resistant to shocks and vibrations, making it suitable for use in manufacturing plants and laboratories where equipment may be subjected to harsh conditions.

Another notable characteristic of the 5725A is its ability to generate stable temperature outputs with minimal drift, contributing to consistency in calibration results. Its built-in self-test functionality helps ensure that the unit remains in peak operating condition, giving users confidence in their calibration processes.

In summary, the Fluke 5725A is a top-tier temperature calibration source that excels in accuracy, reliability, and user-friendliness. Whether in laboratory settings or industrial applications, its advanced technologies and features make it an essential tool for professionals tasked with maintaining precision in temperature measurements.