If you need ZERO calibration but do not need SPAN calibration, turn the power off to exit from the calibration procedure.
7.To calibrate with a different weight, change the displayed value using the
If you enter the wrong number, press the C key. The value returns to the capacity and enter the number again.
8.Place a calibration weight on the platform with the same value as displayed, and wait until the STABLE indicator turns on.
9.Press the PRINT/ENT key. The indicator calibrates
SPAN and end will appear. Remove the weight from the platform, and turn the power off.
-00end00
Note
The value set in step 7 will be cleared after the power is switched off.
If the indicator will be moved to another location, set the gravity acceleration value for the current location and calibrate the indicator according to the procedure above. See the next section to set the value.
7-2. Gravity acceleration correction
When the indicator will be moved to different place from the calibration location, it should be calibrated using a calibration weight.
But if a calibration weight is not available, the gravity acceleration correction will compensate the indicator. To use this function, set the gravity acceleration value to the value of the location where the calibration will be done. Refer to the gravity acceleration map appended to the instruction manual of the
1.At step 3 of the previous section
The integral part “9” is fixed, enter the value after the decimal point.
If you hit the wrong number, press the C key. The value returns to the original number. Enter the number again.
2.Press the PRINT/ENT key, then Cal 0 will be displayed.
3.Go to step 5 of “7-1. Calibration using a weight”.
-00Cal00