A&D 1WMPD4002313 manual Function Parameter Settings, Procedure for setting parameters

Models: 1WMPD4002313

1 24
Download 24 pages 49.96 Kb
Page 12
Image 12

6. FUNCTION PARAMETER SETTINGS

To design a platform scale, set the capacity, weighing unit and other functions according to the specifications of the scale.

Normally, end users do not have to set these function parameters and these setting can be sealed together with the calibration switch.

Parameter

-C01-00

Function number

6-1. The procedure for setting parameters

‰Do not set “C01-1”. Once set, the indicator cannot get into the parameter

setting procedure again.

‰ When turning the power on and the indicator may display ------- ,

press the PRINT/ENT key to proceed to the weighing mode.

‰See “7. CALIBRATION” as for the location of the calibration (CAL) switch.

1.

 

In the weighing mode, press and hold the calibration (CAL)

 

 

switch until

 

Cal

 

 

appears. Then, release the switch.

‰ Press and hold the

 

 

 

key for about 5 seconds, then you

 

0

 

 

can also enter the calibration mode (when “C01-0” is set).

‰ Press the

 

ON/OFF

 

key or (CAL) switch to stop the

 

setting procedure.

 

 

 

 

2.

Press and hold the

 

 

key and press the

 

ZERO

 

 

 

key. Then the first function number is

 

 

PRINT/ENT

 

 

displayed.

 

 

 

 

 

 

 

 

3.Press the 2 key to enter into the “C02 Capacity” setting.

4.Press the PRINT/ENT key. Then the parameter stored appears.

5.Enter the parameter value for the capacity using the 10-key pad.

6.Press the ZERO key to go to the next function number “C03 Weighing display unit”.

7.Set C03 “0” (metric unit), or “1” (non-metric unit).

8.The settings C02 and C03 are minimum requirements to design a platform scale. To finalize the settings, be

sure to press the PRINT/ENTER key. Then, the display shows end for a few seconds and returns to weighing mode.

-00Cal00

-C01-01 -C02-01 -C02-00-C02-02-C03-01-00end00

‰ The ZERO key increments the function number as in the step 6 above, but note that all of the new parameters are not stored until end is displayed.

‰ To exit with nothing changed, turn the power off without pressing the PRINT/ENT key.

‰Set the other parameters according to the specifications required.

10

Page 12
Image 12
A&D 1WMPD4002313 manual Function Parameter Settings, 00Cal00 C01-01 -C02-01 -C02-00-C02-02-C03-01-00end00