MS-70 / MX-50 / MF-50 / ML-50 8
4.
4.4.4. Outline and Outline and Outline and Outline and FeaturesFeaturesFeaturesFeatures The moisture analyzer is built using a super hybrid sensor (S.H .S.) ad opt ed i n an
analytical balance. Therefore, the result is more precise and gets greater repeatability.
An analyzer using the S.H.S. has high sensitivity, needs only a sample quantity of a
few grams, and the analysis time becomes shorter.
A 400W halogen lamp is used as the heating source and the temperature on the
sample pan can reach 200°C within two minutes.
There are five analysis modes.
Standard mode..... The m oisture content can be ob tained wi th setti ngs of th e
drying temperature and accuracy.
Quick mode........... Sample is heated up for approximately three minutes at 200°C so
that analysis time becomes shorter. The moisture content can
be obtained with settings of the drying temperature and
accuracy.
Automatic mode .... When the change of moisture content per one minute is less than
the preset termination value, the measurement is automatically
stopped and the result is o bt aine d.
Timer mode ............ The sample is dried for a preset time and the result is obtained.
Manual mode.......... This mode can stop the measurement by key oper atio n and the
result is decided.
The heating patterns can be used to analysis mode without quick mode.
(For ML-50, standard drying and quick drying can be used only)
Standard drying ..... Maintains a constant drying temperature.
Ramp drying........... Increases the drying temperature gently.
Step drying ............. Uses multiple steps of the drying temperature.
Quick drying........... Heats up 200°C for few minutes and uses a constant drying
temperature.
The analyzer can store and recall proper individual settings for each sample using a
program number (PROG No.).
MS-70 / MX-50 MF-50 ML-50
Maximu m progr am nu m b e r 20 sets 10 sets 5 sets
The data memory function can store results and output them at one time.
MS-70 / MX-50 MF-50 ML-50
Maximum number of storable results 100 50 30
The software "WinCT-Moisture", the standard accessory of the MS-70 and MX-50,
has a function that can make a graph of the change of moisture content in realtime
and has an optimum temperature search program that judges heating at an
appropriate temperature setting.
The software " WinCT", the standard accessory of the MF-50, is communication
software for transmitting data to a computer using Microsoft Windows.