MS-70 / MX-50 / MF-50 / ML-50 38
Set the Minimum Scale Value of the % Display
Set the Minimum Scale Value of the % DisplaySet the Minimum Scale Value of the % Display
Set the Minimum Scale Value of the % Display
13. Select the % display wit h t he SELECT key.
Select 0.01 [%] with the or key.
Set the Minimum Scale Value of the Gram Display
Set the Minimum Scale Value of the Gram DisplaySet the Minimum Scale Value of the Gram Display
Set the Minimum Scale Value of the Gram Display
14. Select the gra m displ ay with the SELECT key.
Select 0.001 [g] with the or key.
Store the Parameters and Finish the Operation
Store the Parameters and Finish the OperationStore the Parameters and Finish the Operation
Store the Parameters and Finish the Operation
15. Press the ENTER key to store the new parameters of
the measurement program to program number 3.
Pressing the key, the weighing mode is automatically
displayed.
When PROG 3 is recalled, the setti n gs can be used.
To cancel new parameters and return to weighing
mode, press the RESET key.
Timer Mode
Timer ModeTimer Mode
Timer Mode
Refer to page 36 for Standard Mode, Automatic Mode or Manual Mode
This explanation uses the following parameters and sample displays of MX-50.
"Temperature 1" can be set higher than "temperature 2" in step drying.
Common Items
Program number.......................................... 6 ( PROG 6 )
Analysis mode ............................................. Timer mode (Symbol: U-t )
Drying program Heating pattern .................... Step dryi ng ( )
Drying temperature 1 ..........160°C
Drying temperature 2 ..........120°C
Time 1 ..................................5.0 minute s
Time 2 ..................................10. 0 minut es
Measurement unit ............................................... Moistur e con tent
Minimum scale value during measurement..0.01 %
Minimum scale value of the gram display.... 0.001 g
Sample quantity........................................... Approximately 5 g
Data memory function.................................. Not used
Time 1
Drying temperature 2
Drying temperature 1
Stage 1 Stage 2
Time 2
% MOIST
D /W /D
0.000g
1
6
0°C
Gram display of
weighing mode
"0."0"1%
Minimum scale
value of % display
"0."0"0"1g
Minimum scale
value of gram
display