MS-70 / MX-50 / MF-50 / ML-50 35
Select an Analysis Mode
Select an Analysis ModeSelect an Analysis Mode
Select an Analysis Mode
5. Press the SELECT key to select a symbol, it
will blink. Select U-t of the timer mode with
the or key. (Press these keys to select it)
Set the Heating Pattern
Set the Heating PatternSet the Heating Pattern
Set the Heating Pattern
6. Select drying temperature with the SELECT key.
Select ramp drying of th e heating pattern
with the PROGRAM key.
Set the Final Drying Temperature
Set the Final Drying TemperatureSet the Final Drying Temperature
Set the Final Drying Temperature
7. Set 160°C with the or ke y.
Set the Ramp Time
Set the Ramp TimeSet the Ramp Time
Set the Ramp Time
8. Set 5.0 minutes with the or key.
Set the Analysis Time
Set the Analysis TimeSet the Analysis Time
Set the Analysis Time
9. Set 10.0 minutes with the or ke y.
Set a Measurement Unit
Set a Measurement UnitSet a Measurement Unit
Set a Measurement Unit
10. Select a measurement unit with the SELECT
key. Select moisture content (based on a wet
sample ) with the or key.
Set the Minimum Scale Value of the % Display
Set the Minimum Scale Value of the % DisplaySet the Minimum Scale Value of the % Display
Set the Minimum Scale Value of the % Display
11. Select the % d isplay with the SELECT key.
Select 0.01 [%] with the or key.
Set the Minimum Scale Value of the G
Set the Minimum Scale Value of the GSet the Minimum Scale Value of the G
Set the Minimum Scale Value of the Gram Display
ram Displayram Display
ram Display
12. Select g ram display with the SELECT key.
Select 0.001 [g] with the or key.
Store the Parameters and Finish the Operation
Store the Parameters and Finish the OperationStore the Parameters and Finish the Operation
Store the Parameters and Finish the Operation
13. Press the ENTER key to store the new parameters
of the measurement program to program number 4.
Pressing the key, the weighing mode is
automatically displayed. When PROG 4 is recalled,
the settings can be used.
To cancel the new parameters and return to the
weighing mode, press the RESET key.
"U"-"t 0.000g
Analysis mode
1
0.000g
"1"0"5°C
Ramp Time
"5."0 min
2
0"%
"M"O"I"S"T
"/"W
Measurement unit
"0."0"1%
Minimum scale
value of % display
"0."0"0"1g
Minimum scale
value of gram
display
0.000g
1
6
0°C
10.0 min
"1"0."0 min
Analysis time
0.000g
"1"6"0°C
Final drying temperature
Ramp drying
Gram display of
weighing mode