MS-70 / MX-50 / MF-50 / ML-50 29
8.2.
8.2.8.2.8.2. Procedures to Store a Measurement ProgramProcedures to Store a Measurement ProgramProcedures to Store a Measurement ProgramProcedures to Store a Measurement Program

8.2.1.

8.2.1.8.2.1.

8.2.1. Standard Drying

Standard DryingStandard Drying

Standard Drying

This explanation uses the following parameters and sample displays of MX-50.
Common Items
Program number.......................................... 2 ( PROG 2 )
Drying program Heating pattern .................... Standard drying ( )
Drying temperature..............160°C
Measurement unit ............................................... Moistur e con tent
Minimum scale value during measurement..0.01 %
Minimum scale value of gram display.......... 0.001 g
Sample quantity........................................... Approximately 5 g
Data memory function.................................. Not used
Items for Standard Mode
Analysis mode ............................................. Standard mode (Symbol: 5td )
ACCURACY................................................. MID.
Analyzing mode to finish a measurement ......... Automatic setting by ACCURACY
Items for Automatic Mode
Analysis mode ............................................. Automatic mode (Symbol: U-a )
Analyzing mode to finish measurement ............ 0.05 % /min
Items for Timer Mode
Analysis mode ............................................. Timer mode (Symbol: U-t )
Analyzing mode to finish measurement ............ 10 min utes
Items for Manual Mode
Analysis mode ............................................. Manual mode (Symbol: U-m )
1. Display the gram unit (of the weighing mode).
Select a Program Number to Edit the Measurement Program
Select a Program Number to Edit the Measurement ProgramSelect a Program Number to Edit the Measurement Program
Select a Program Number to Edit the Measurement Program
2. Press the PROGRAM key and press the or
key to select a program number.
3. Press the ENTER key to use the number.
4. The analyzer displays end and returns to the
weighing mode.
Caution If the data memory function is active, the data
memory number (MEM) is displayed in place of the
program number (PROG).
Drying temperature
Time
Drying temperature
Standard Mode, Automatic Mode, Manual Mode Timer Mode
% MOIST
D /W /D
PROG 1 0.000g
PROG "2
end
PROG 2 0.000g
Program number