2 - Verification and Performance Tests

Table 2-1. Test Equipment Required for Verification and Performance Tests

Type

Specifications

Recommended Model

Current Monitor Resistor

15 A (0.1 ohm) 0.04%

Guildline 9230/15

DC Power Supply

Minimum 5 A output current rating

Agilent 6632B

Digital Voltmeter

Resolution: 10 nV @ 1V

Agilent

3458A or equivalent

 

Readout: 8 1/2 digits

 

 

 

Accuracy: 20 ppm

 

 

Electronic Load

100V, 5 A minimum, with transient capability

Agilent

6060B (60V max.), 6063B

 

 

(240V) or equivalent

GPIB Controller

Controller with full GPIB capabilities

HP Series 300 or equivalent

Resistors

400 ohm, 5W

Agilent p/n 0811-1857

 

1 ohm, 100 W (or 2 ohm adjustable)

Ohmite D12K2R0 (2 ohm adjustable)

(Load resistors may

0.6 ohm, 100W (6611C)

 

 

substitute for electronic

9 ohm, 100W (6612C)

 

 

load if load is too noisy

49 ohm, 100W (6613C)

 

 

for CC PARD test)

 

 

99 ohm, 100W (6614C)

 

 

 

 

 

 

or an appropriate 150W Rheostat

 

 

Oscilloscope

Sensitivity: 1 mV

Agilent

54504A or equivalent

 

Bandwidth Limit: 20 MHz

 

 

 

Probe: 1:1 with RF tip

 

 

RMS Voltmeter

True RMS

Agilent 3400B or equivalent

 

Bandwidth: 20 MHz

 

 

 

Sensitivity: 100 ∝ V

 

 

Variable-Voltage

Adjustable to highest rated input voltage

 

 

Transformer

range.

 

 

 

Power: 500 VA

 

 

Measurement Techniques

Test Setup

All tests are performed at the rear terminals of the supply as shown in Figure 2-1. Measure the dc voltage directly at the +S and -S terminals. Set the Remote/Local switch to Remote and connect the output for remote sensing. Use adequate wire gauge for the load leads.