2 - Verification and Performance Tests
Table
Type | Specifications | Recommended Model | |
Current Monitor Resistor | 15 A (0.1 ohm) 0.04% | Guildline 9230/15 | |
DC Power Supply | Minimum 5 A output current rating | Agilent 6632B | |
Digital Voltmeter | Resolution: 10 nV @ 1V | Agilent | 3458A or equivalent |
| Readout: 8 1/2 digits |
|
|
| Accuracy: 20 ppm |
|
|
Electronic Load | 100V, 5 A minimum, with transient capability | Agilent | 6060B (60V max.), 6063B |
|
| (240V) or equivalent | |
GPIB Controller | Controller with full GPIB capabilities | HP Series 300 or equivalent | |
Resistors | 400 ohm, 5W | Agilent p/n | |
| 1 ohm, 100 W (or 2 ohm adjustable) | Ohmite D12K2R0 (2 ohm adjustable) | |
(Load resistors may | 0.6 ohm, 100W (6611C) |
|
|
substitute for electronic | 9 ohm, 100W (6612C) |
|
|
load if load is too noisy | 49 ohm, 100W (6613C) |
|
|
for CC PARD test) |
|
| |
99 ohm, 100W (6614C) |
|
| |
|
|
| |
| or an appropriate 150W Rheostat |
|
|
Oscilloscope | Sensitivity: 1 mV | Agilent | 54504A or equivalent |
| Bandwidth Limit: 20 MHz |
|
|
| Probe: 1:1 with RF tip |
|
|
RMS Voltmeter | True RMS | Agilent 3400B or equivalent | |
| Bandwidth: 20 MHz |
|
|
| Sensitivity: 100 ∝ V |
|
|
Adjustable to highest rated input voltage |
|
| |
Transformer | range. |
|
|
| Power: 500 VA |
|
|
Measurement Techniques
Test Setup
All tests are performed at the rear terminals of the supply as shown in Figure
12