Troubleshooting - 3
Specific Troubleshooting Procedures
Power-on Self-test Failures
The
NOTE: A partial self test is performed when the *TST? query is executed. Those tests that interfere with normal interface operation or cause the output to change are not performed by *TST?. The return value of *TST? will be zero if all tests pass, or the error code of the first test that failed. The power supply will continue normal operation if *TST? returns a
Table 3-2. Self-Test Error Codes/Messages
Error Code | Description | Probable Cause | |
E1 | Checksum in | A2 Interface Bd | |
E2 | Checksum in Config | A2 Interface Bd | |
E3 | Checksum in Cal | A2 Interface Bd | |
E4 | Checksum in State | A2 Interface Bd | |
E5 | Checksum in RST | A2 Interface Bd | |
E10 | RAM test failed | A2 Interface Bd | |
E11 | 12 bit DAC test failed, 0 is written to DAC U241A and B, | A2 | Interface Bd |
| ADC U242 is checked for 133 +/- 7 counts |
|
|
E12 | 12 bit DAC test failed, 4095 is written to DAC U241A | A2 | Interface Bd |
| and 0 to B, ADC U242 is checked for 71 +/- 7 counts |
|
|
E13 | 12 bit DAC test failed, 0 is written to DAC U241A and | A2 | Interface Bd |
| 4095 to B, ADC U242 is checked for 71 +/- 7 counts |
|
|
E14 | 12 bit DAC test failed, 4095 is written to DAC U241A | A2 | Interface Bd |
| and B, ADC U242 is checked for 10 +/- 7 counts |
|
|
E15 | 8 bit DAC test failed, 10 and 240 are written to DAC | A2 | Interface Bd |
| U244, ADC U242 is checked for 10 and 240 +/- 7 counts |
|
|
E80 | Dig I/O test failed, SEC_PCLR written low and high, | A2 | Interface Bd |
| read back through Xilinx |
|
|
E213 | A2 | Interface Bd | |
E216 | A2 | Interface Bd | |
E217 | A2 | Interface Bd | |
E218 | A2 | Interface Bd | |
E220 | Front Panel comm UART input overrun | A3 | Front Panel/Display Bd |
33