Agilent Technologies 5962-8200 service manual Principles of Operation

Models: 5962-8200

1 58
Download 58 pages 2.56 Kb
Page 47
Image 47

Principles of Operation - 4

When the downprogramming stage is turned on (in either CV or CC mode), the CV/CC control circuit causes the Control signal to go low, which in turn causes the downprogramming transistors to conduct current away from the load and speed up downprogramming.

During operation, a PM_Inhibit signal will cause the output stage bias/shutdown circuit to turn off the gated

15 V bias voltages and shut down the output if any of the following occur: The output is programmed off.

An overvoltage condition is detected (OV_Detect* signal is received). The line voltage falls below 90 volts (approximately).

Current readback is provided by three separate circuits. The previously discussed high range current signal (Imon_H) returns the high range currrent measurement. When the unit is operating in the low current readback mode, a separate low range current shunt and amplifier provides low-current readback via the Imon_L signal . A shunt clamp (Q302 and Q304) clamps the voltage across RmLo to approximately 1.8 V. The third current readback circuit consists of a high bandwidth current amplifier that returns dynamic current measurements from the output filter capacitor via the Imon_P signal. Note that the Imon_H and the Imon_P signals are combined to return the actual output current measurement.

An overvoltage detect circuit compares the output voltage to the programmed overvoltage setting. When the output exceeds the programmed setting, the OV_Detect* signal goes low, which informs the logic array that an OV condition has occurred. The crowbar control circuit is enabled when the OV_SCR* signal is received. When an overvoltage condition occurs, the SCR control circuit generates the OV signal, which causes the following actions to occur:

1.The SCR fires, shorting the supply's output.

2.The microprocessor circuits are notified of the OV condition (OV_Detect* is low) in order to program the ouput off, turn off the gated 15V bias supplies, and update the status of the unit.

3.The PM_Inhibit signal goes high, programming the output off and shutting down the gated 15V bias for the output regulators.

4.When a output protection clear command is executed, the microprocessor circuits resets the OV circuits, turns on the gated 15V biases, and programs the output to its previous level.

The fan driver control circuit provides the DC voltage to operate the cooling fan. The Fan_Prog signal from the secondary interface circuit varies this voltage according to the ambient and heatsink temperature as well as the output voltage and current of the supply.

47

Page 47
Image 47
Agilent Technologies 5962-8200 service manual Principles of Operation