Troubleshooting - 3

J307 Voltage Measurements

Cable W9 connects J307 of the A1 Main Board Assembly to J207 of the A2 Interface Assembly. Table 3-4 provides a quick method of determining if the voltages between these assemblies are within the normal range. If any of these voltages is outside the normal range, refer to the flowcharts to further troubleshoot the circuit associated with the abnormal voltage.

Table 3-4. Voltage Measurements at J207 (A2 Interface to A1 Main board)

A1J207

Signal Name

CV Mode

CC Mode

Pin #

 

Full Scale Voltage

Full Scale Voltage

 

 

No Load

Full Load

1

PM_INHIBIT (Enabled)

0

0

2

OV_SCR*

+5

+5

3

OV_PROG

+3.9

+3.9

4

FAN_PROG

+2.8

+3.8

5

OV_DETECT*

+5

+5

6

SW_POS (Norm)

+5

+5

7

RANGE_SELECT (High)

0

0

8

OS_TRIM_NEG (COMP)

+1.7

+1.7

 

OS_TRIM_NEG (SCPI)

+4.0

+4.0

9

+5Vs

+5

+5

10

COMMON

0

0

11

COMMON

0

0

12

+15Vs

+15

+15

13

-15Vs

-15

-15

14

HS_THERM (@25C)

+2.5

+2.5

15

FUSE

+2.4

+2.6

16

IMON_H

0

+3.5

17

IMON_L

0

+14.7

 

IMON_L (@20mA Out)

+4.8

+4.8

18

IMON_P

0

0

19

VMON

+4.8

+4.8

20

COMMON

0

0

21

COMMON

0

0

22

COMMON

0

0

23

COMMON

0

0

24

CV_PROG

-4.8

-4.8

25

CC_PROG

-4.8

-4.8

26

CC_DETECT*

+5

0

27

CCN_DETECT*

+5

+5

28

CV_DETECT*

0

+5

35

Page 35
Image 35
Agilent Technologies 5962-8200 J307 Voltage Measurements, Voltage Measurements at J207 A2 Interface to A1 Main board