Troubleshooting - 3

From Sheet 2

Program current to full scale, voltage to Vmax and load to the power supply's rated current. Supply should be in CC.

Will not go into CC or error > 10%

?

No

Output out of spec but close?

No

Output OK but meter wrong?

No

 

CC_Prog OK ?

 

 

 

Replace A1

 

 

Yes

 

Yes

(see Table 3-4)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

No

Replace A2

 

 

 

Calibrate unit

Yes

 

 

 

 

 

 

 

 

 

Calibrate current. If still

 

 

wrong or will not

Yes

 

 

calibrate, replace A2

 

 

 

Turn on OCP and insure Protect trips.

Prot trips?

 

 

 

CC_detect* low?

 

 

 

 

Check cable W9,

 

No

 

No

 

 

 

(see Table 3-4)

 

 

replace A1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Yes

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Replace A2

 

 

 

 

 

Yes

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Goto Sheet 4

Figure 3-1 Sheet 3. Troubleshooting Flowchart

31

Page 31
Image 31
Agilent Technologies 5962-8200 service manual Sheet 3. Troubleshooting Flowchart