3 - Troubleshooting

From Sheet 3

Connect controller to the

HPIB port and send commands to set the output voltage and current and readback the output.

Accepts and reads

 

No

 

back?

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Yes

 

 

 

Replace A2

Run the Performance

Test in Chapter 2.

Passes test?

No

Yes

Regulation, Transient Response and ripple problems are generally caused by A1

Short RI terminals on rear of supply and insure output disables and Prot annunciator comes on.

Remote Inhibit

No

OK?

 

Yes

There is either no fault with the power supply or the problem is not covered by this procedure.

Replace A2

Figure 3-1 Sheet 4. Troubleshooting Flowchart

32

Page 32
Image 32
Agilent Technologies 5962-8200 service manual Sheet 4. Troubleshooting Flowchart