Troubleshooting - 3

Turn on unit and observe the display. All of the segments and annunciators, the address and then after self test should display an error message or go to the metering mode.

Check Bias voltages (see Table 3-3)

Bias voltages OK?

 

 

 

Transformer Inputs

 

 

 

Check Main Fuse,

 

No

 

 

No

 

 

 

 

 

Replace T1

 

 

 

 

 

OK?

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Yes

 

 

 

 

 

Yes

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Replace A1

 

Display comes on?

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

No

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Yes

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Error Message?

 

 

 

Go to Error Message

 

 

Yes

 

 

 

 

Table 3-2.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

No

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Check A1F305,

+5V @ A2J211-1 (to

 

 

 

Red/White/Black

 

No

 

cable A1-A2 &

 

chassis)?

 

 

 

cable A2-A3, track

 

 

 

 

 

 

 

on A2 (J206-J211)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Yes

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

A3J111-5 low (no

 

 

 

Replace A3

 

 

No

 

 

 

pulses)?

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Yes

 

 

 

 

 

 

Protect

annunciator Yes RI? on?

Yes Replace A2

No

No

OV?

No

OC?

No

FS?

No

 

 

Check for OV setting <

 

 

Voltage setting, Replace

Yes

 

 

A1

 

 

 

 

 

 

 

 

 

Check that OCP is not

 

Yes

 

 

 

enabled, Replace A1

 

 

 

 

 

 

 

 

 

 

 

 

 

Check F309 (fuse near

 

Yes

 

main heat sink),

 

 

 

Replace A1

 

 

 

 

 

Go to Sheet 2

For OT check fan, Replace A1

Figure 3-1 Sheet 1. Troubleshooting Flowchart

29

Page 29
Image 29
Agilent Technologies 5962-8200 service manual Sheet 1. Troubleshooting Flowchart