3 - Troubleshooting

From Sheet 1

Enable output and program voltage and current full scale with no load. Measure output voltage.

 

 

 

 

 

Check to insure OV

Unit OV's?

 

 

 

setting is not less than

 

Yes

 

 

 

the voltage setting. If not

 

 

 

 

 

 

 

 

then replace A1.

 

 

 

 

 

 

 

 

 

 

 

 

No

Output voltage

 

 

 

Yes

> 10% error?

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

No

 

 

 

CV_Prog &

 

 

 

Check cable W9,

CC_Prog OK? (see

 

 

 

 

Yes

 

 

 

Replace A1

Table 3-4)

 

 

 

 

 

 

 

No

Replace A2

Output out of spec

 

 

 

Calibrate voltage

 

 

Yes

 

 

 

but close?

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

No

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Output OK but

 

 

 

Calibrate voltage. If

 

 

Yes

 

still wrong or will not

 

meter wrong?

 

 

 

 

 

 

calibrate, replace A2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

No

 

 

 

 

 

 

Program the OV 2 volts lower than the output voltage.

Unit OV's? No Program OV to full scale

Yes

OV_Prog OK?

 

 

 

Calibrate OV. If OV is still

 

Yes

 

not functioning properly

(see Table 3-4)

 

 

 

 

 

check W9, replace A1.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

No

 

 

 

 

 

Go to Sheet 3

Replace A2

Figure 3-1 Sheet 2. Troubleshooting Flowchart

30

Page 30
Image 30
Agilent Technologies 5962-8200 service manual Sheet 2. Troubleshooting Flowchart