3 - Troubleshooting
From Sheet 1
Enable output and program voltage and current full scale with no load. Measure output voltage.
|
|
|
|
| Check to insure OV |
Unit OV's? |
|
|
| setting is not less than | |
| Yes |
| |||
|
| the voltage setting. If not | |||
|
|
| |||
|
|
|
|
| then replace A1. |
|
|
|
|
|
|
|
|
|
|
|
|
No
Output voltage |
|
| |||
| Yes | ||||
> 10% error? |
| ||||
|
| ||||
|
|
|
|
| |
|
|
|
|
|
|
| No |
|
|
|
CV_Prog & |
|
|
| Check cable W9, |
CC_Prog OK? (see |
|
|
| |
| Yes |
| ||
|
| Replace A1 | ||
Table |
|
|
| |
|
|
|
|
No
Replace A2
Output out of spec |
|
|
| Calibrate voltage |
| |||
| Yes |
| ||||||
|
| |||||||
but close? |
|
|
|
|
| |||
|
|
|
|
| ||||
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| No |
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
Output OK but |
|
|
| Calibrate voltage. If |
| |||
| Yes |
| still wrong or will not |
| ||||
meter wrong? |
|
|
| |||||
|
|
| calibrate, replace A2 |
| ||||
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
| ||
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
| |
| No |
|
|
|
|
|
|
Program the OV 2 volts lower than the output voltage.
Unit OV's? No Program OV to full scale
Yes
OV_Prog OK? |
|
|
| Calibrate OV. If OV is still | |||
| Yes |
| not functioning properly | ||||
(see Table |
|
| |||||
|
|
| check W9, replace A1. | ||||
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| No |
|
|
|
|
|
Go to Sheet 3
Replace A2
Figure 3-1 Sheet 2. Troubleshooting Flowchart
30