Safety Requirements Power Requirements
10 Agilent 81133A/81134A Pulse Generator User’s Guide, January 2005
Power Requirements
NOTE When the front panel switch i s off, the instrument is in “stan dby”
mode. The instrument is disconnected from the AC line power only by
disconnecting the power cord.
The instrument can operate from any single-phase AC power source
supplying 100 – 240 V in the frequency range from 47 – 63 Hz. The
maximum power consumption is 200 VA with all options installed.
When the instrument is switched on the power supply adapts
automatically to the applied AC power (Auto Selecti on) and monitors
the AC power range during operation.
Performance Requirements
For best performance of data transfer and for EMC compliance use an
Agilent 82357-61601 Mini-USB cable only.
Ventilation Requirements
Make sure that there is adequate clearance of 50 mm (2 in) at the top
and right side to ensure adequ ate air flow. If the air flow is restrict ed,
the internal operating temperatu re will be higher, reducing the
instrument's reliability.
NOTE Do not cover the ventilation holes.