Introduction to the Agilent 81133A/81134A Pulse Generator Front Panel Overview
16 Agilent 81133A/81134A Pulse Generator User’s Guide, January 2005
Input and Outputs
The Agilent 81133A/81134A pulse generator provides the following
inputs and outputs:
Channel Specific Inputs and O utputs For each channel:
•One Delay Control Input to apply an external signal for jitter
modulation.
For more information, see “Jitter Modulation” on page 83.
One normal and one inverted Output for the generated signal
Global Inputs and Output s For the i nstrume nt:
•Clock Input
The clock input is used to apply an external clock signal or a
reference signal for the internal PLL if a higher frequency accuracy
than 50ppm is required.
This inpu t can be AC or DC ter minated . If it is D C is ter minated, the
termination voltage can be set.
For more information, see “Clock Sources” on page 73.
•Start Input
The st art i nput is used t o star t the instr ument on an e xtern al sig nal.
For more information, see “Start Mode” on page 75.
Trigger Output
The trigger output can be used to trigger another instrument.
For more information, see “Trigger Out” on page 75.
Global Inputs
and Outputs
Channel Inputs
and Outputs