Introduction to the Agilent 81133A/81134A Pulse Generator Front Panel Overview
16 Agilent 81133A/81134A Pulse Generator User’s Guide, January 2005
Input and OutputsThe Agilent 81133A/81134A pulse generator provides the following
inputs and outputs:
Channel Specific Inputs and O utputs For each channel:
•One Delay Control Input to apply an external signal for jitter
modulation.
For more information, see “Jitter Modulation” on page 83.
• One normal and one inverted Output for the generated signal
Global Inputs and Output s For the i nstrume nt:
•Clock Input
The clock input is used to apply an external clock signal or a
reference signal for the internal PLL if a higher frequency accuracy
than 50ppm is required.
This inpu t can be AC or DC ter minated . If it is D C is ter minated, the
termination voltage can be set.
For more information, see “Clock Sources” on page 73.
•Start Input
The st art i nput is used t o star t the instr ument on an e xtern al sig nal.
For more information, see “Start Mode” on page 75.
• Trigger Output
The trigger output can be used to trigger another instrument.
For more information, see “Trigger Out” on page 75.
Global Inputs
and Outputs
Channel Inputs
and Outputs