Setting Up a Burst Signal Setting Up Generic and Advanced Signals
Agilent 81133A/81134A Pulse Generator User’s Guide, January 2005 49
Reconnect the DUT Enable the outputs:
Auxiliary Settings for Start Input and
Trigger Output
To set the start input and trigger output parameters:
1Switch to the Aux page.
2In the Start Input panel, set:
–The start mode to Ext. Inpu t.
This automatically puts the instrument in “armed” mode. This is
indicated by the activated Armed softkey function.
“Armed” mode means that the signal is output as soon as the
specified external signal occurs at the start input.
NOTE In burst mode, the “armed” mode is deac tivated as soon as the
specified burst has been ou tput.
To put the instrument in “armed ” mode again, press the Armed
softkey function.
Start on the Rising edge.
The threshold t o 500 mV.
3In the Trigger Output panel:
Select the trigger m ode Pulse.
The frequency of the trigger output equa ls the frequency as
specified in the instru ment panel.
Set Ampl to 2 V.
Set Offset to 0 V.
Enable the trigger output.