Setting Up Generic and Advanced Signals Setting Up a Repeti tive Burst Signal
52 Agilent 81133A/81134A Pulse Generator User’s Guide, January 2005
Setting Up a Repetitive Burst Signal
Task S et up a signal with 4 repeated bursts.
Each burst consists of a 32-bit pattern signal with NRZ data output
format at 100 MHz to check a digital circuit with LVPECL logic.
The data pattern is:
11100000000000000000000000000000
Using the Graphical User Interface
Instrument Settings To set the instrument parameters:
1Disable the outputs.
2In the instrument panel, choose the RBurst mode.
3Enter 4 for the number of null data packets.
4Define that the number of zeros t hat follows the repeated bursts is
as large as the burst length.
5Set the frequency to 100 MHz.
Data Data Data Data Data Data Data
Data Length = 32 x 10 ns
Burst Length = 4 x Data Length Number of Null data packets =
Burst Length