Setting Up a Repetitive Burst Signal Setting Up Generic and Advanced Signals
Agilent 81133A/81134A Pulse Generator User’s Guide, January 2005 53
The instrument panel now looks as follows:
Channel Settings for the PRBS Signal To set the channel paramete rs:
1Switch to the Channel page.
2In the Timing panel:
Set the signal mode to NRZ.
–Set the Dela y to 50 ps.
3In the Levels panel, set:
The level format to Custom.
Amplitude to 2.0 V.
Offset to 0.0 V.
4Enable the Channel 1 normal Ou tput:
By pressing the respective sof tkey in the user interface:
– OR –
By press ing the nor mal Out so ftkey for cha nnel 1 next to the Out put
port.
Burst Repetition
Instrument
Mode
Specifies the
number of zeros