Using the Agilent 81133A/81134A Pulse Generator Timing of Generated Signals
78 Agilent 81133A/81134A Pulse Generator User’s Guide, January 2005
Pulses
The following figure shows the typical timing for contin uous pulses.
Characteristics • Puls e periods can be either generated continuously or can be
started manually or by the a rming source.
Delay, pulse width (or duty cycle) and deskew can be set for each
channel.
• The instrument mode is Pulse/Pattern.
• The pattern mode is Pulse.
• For the clock source, you can select from:
Internal (YIG Oscillator)
External signal at Clock Input
External 10 MHz Reference at Clock Input
For more information about the clock sources, see “Clock Sources”
on page 73.
The Trigger Out is generated with every clock pu lse, but can be
optionally divided by any number in the range 1 ... 231 – 1.
See also “Trigger Out” on page 75.
Output
Channel 1
Output
Channel 2
Freq. Divider = 2
Duty Cycle
Delay
Trigger Out