Use, Maintenance, and Care of the Devices

Gaging Connectors

If the pin depth is outside the observed limits listed in Table 2-2 on page 2-4, use the following procedure to reset it. Always measure the sliding load pin depth before attaching it to any connector.

This procedure assumes that you were directed here from “Gaging the Sliding Load” on page 3-12. If not, perform the steps in that procedure before performing this procedure.

1.The gage should be attached to the sliding load. The sliding load should have its centering bead installed. Refer to “Gaging the Sliding Load” on page 3-12if necessary.

2.With a 0.050 inch hex key, loosen the two largest hex screws by turning them 1/4 turn clockwise. Refer to Figure 3-6.

CAUTION Do not loosen any hex screws other than the two largest hex screws pointed out in Figure 3-6.

3.Gently turn the center conductor pin depth adjustment knob on the sliding load until the gage pointer reads −3.81 micrometers (−1.5 x 10–4inches). Refer to Figure 3-6.

4.Tighten the two hex screws just until they are finger tight (do not overtighten).

5.Wait approximately five minutes to allow the temperature to stabilize. Do not touch either the gage or the sliding load during this time.

6.Note the gage reading. If it is no longer within the allowable range, perform steps 2–5 again.

7.Loosen the connecting nut and remove the gage from the sliding load. If you’re going to store the sliding load, leave the centering bead installed. If you’re gong to use the sliding load for an electrical calibration, remove the centering bead.

NOTE

When performing pin depth measurements, use different orientations of the

 

gage within the connector. Averaging a minimum of three readings, each

 

taken after a quarter-turn rotation of the gage, reduces measurement

 

variations that result from the gage or the connector face not being exactly

 

perpendicular to the center axis.

 

 

The sliding load pin depth is now is specification and the load is ready to use. Once the sliding load pin depth is set it rarely needs to be adjusted. However, the pin depth should be rechecked before each use. Replace the protective plastic caps on the sliding load and gage connectors when these devices are not in use.

3-14

85054B

Page 32
Image 32
Agilent Technologies 85054-90049 manual Gaging Connectors

85054-90049 specifications

Agilent Technologies 85054-90049 is a high-performance calibration kit designed for vector network analysis across a wide range of frequencies. This kit is an indispensable tool for engineers and technicians involved in the measurement and characterization of RF components, ensuring accuracy and reliability in their testing processes.

One of the main features of the 85054-90049 calibration kit is its ability to provide accurate calibration across multiple frequency bands, including microwave and millimeter-wave frequencies. This versatility is crucial for professionals who work with different types of RF devices, such as amplifiers, filters, and antennas. The kit supports a frequency range of up to 67 GHz, making it suitable for a variety of applications in the telecommunications and aerospace industries.

The calibration kit utilizes precision standards and adapters that ensure low insertion loss and minimal reflection. These components are engineered to provide high stability and repeatability in measurements, which is vital for maintaining quality control in manufacturing processes or during research and development stages. The accuracy of the calibration is further enhanced by the materials used in the construction of the connectors and adapters, which are designed to minimize signal degradation.

Another key technology employed in the Agilent 85054-90049 is its rugged design, intended to withstand the demands of laboratory environments as well as field applications. The kit is portable, allowing engineers to perform precision measurements wherever needed without compromising on performance. Its compact design and durable case make it an ideal choice for field applications.

The 85054-90049 calibration kit is also compatible with Agilent's range of vector network analyzers, creating a seamless user experience. The calibration process, supported by comprehensive software tools, is designed to be user-friendly, allowing technicians to perform calibrations quickly and accurately. In addition, the kit comes with detailed documentation that includes calibration procedures, ensuring that even those new to RF testing can follow along with ease.

In summary, the Agilent Technologies 85054-90049 is a state-of-the-art calibration kit that combines high performance, compatibility, and durability. Its advanced features cater to various RF testing needs, making it a vital resource for professionals in industries where precision measurements are of utmost importance. Whether in laboratories or field settings, this calibration kit stands out as a benchmark for quality in RF and microwave measurement applications.