Frequency characteristics

 

Range

30 kHz to 3 GHz

Option 006

30 kHz to 6 GHz

Resolution

1 Hz

Stability9

±7.5 ppm (0° to 55° C)

 

±3 ppm/year

Option 1D5

±0.05 ppm (0° to 55° C),

 

±0.5 ppm/year

Accuracy

±10 ppm at 25° ±5° C

Power range2

-85 to +10 dBm

Option 075/014

-85 to +8 dBm

Power sweep range

25 dB (typically 31 dB)

Resolution

0.01 dB

Level accuracy1,2,5

±1.0 dB

Level linearity1,2,5

-15 dBm to +5 dBm ±0.2 dB

 

5 dBm to 10 dBm6 ±0.5 dB

Impedance9

50 Ω (nominal)

30 kHz to 3 GHz

>16 dB RL (<1.38 SWR)

3 GHz to 6 GHz

>14 dB RL (<1.50 SWR)

Option 075

75 Ω (nominal)

30 kHz to 3 GHz

>16 dB RL (<1.38 SWR)

Spectral purity

 

2nd harmonic3

<-40 dBc at 0 dBm9

 

<-50 dBc at -10 dBm9

Option 002

<-25 dBc at 10 dBm6

3rd harmonic4

<-40 dBc at 0 dBm9

 

<-50 dBc at -10 dBm9

Option 002

<-25 dBc at 10 dBm6

Nonharmonic spurious9

 

Mixer related

<-30 dBc at 10 dBm6

 

<-55 dBc at -10 dBm

Test-Port Input Characteristics

Frequency range

30 kHz to 3 GHz

Option 006

30 kHz to 6 GHz

Average noise level2,8

 

<3 GHz

<-82 dBm (3 kHz BW)

<3 GHz

<-102 dBm (10 Hz BW)

<3 GHz

<-110 dBm (10 Hz BW) typical

3 to 6 GHz

<-77 dBm (3 kHz BW)

3 to 6 GHz

<-97 dBm (10 Hz BW)

3 to 6 GHz

<-105 dBm (10 Hz BW) typical

Maximum input level

10 dBm

Damage level

26 dBm or 35 VDC

Impedance

50 Ω (nominal)

Option 075

75 Ω (nominal)

Frequency response2,5

 

300 kHz to 3 GHz

±1.0 dB

3 GHz to 6 GHz

±2.0 dB

Internally generated harmonics (Option 002)

2nd harmonic3

<-15 dBc at +8 dBm

 

<-30 dBc at 0 dBm9

 

<-45 dBc at -15 dBm9

3rd harmonic4

<-30 dBc at +8 dBm

 

<-50 dBc at 0 dBm9

 

<-50 dBc at -15 dBm9

Harmonic measurement accuracy10

300 kHz to 3 GHz

±1.5 dB

3 GHz to 6 GHz

±3 dB (Option 006)

Harmonic measurement dynamic range

-40 dBc (output = -10 dBm, input = <-15 dBm)9

Frequency offset mode11

 

Frequency range

300 kHz to 3 GHz

Option 006

300 kHz to 6 GHz

R-channel input requirements

 

Power level

 

300 kHz to 3 GHz

0 to -35 dBm

3 GHz to 6 GHz

0 to -30 dBm

LO spectral purity (typical)

 

Maximum spurious input

<-25 dBc

Residual FM

<20 kHz

LO frequency accuracy9

±1 MHz at nominal frequency

1.

Relative to 0 dBm output power.

2.

Typical below 300 kHz.

3.

16 MHz to 3 GHz.

4.

16 MHz to 2 GHz.

5.

Typical from 2 to 3 GHz for instruments with Option 075.

6.

+8 dBm maximum with Option 075, or Option 014.

7.

Test performed on port 1 only.

8.

Instruments with Option 075 are degraded 2 dB.

9.

Typical performance.

10.

25° ±5° C.

11. The Agilent 8753ES source characteristics and measurement accuracy in this

 

mode are dependent on the stability of the external LO source. The RF source

 

tracks the LO to maintain a stable IF signal at the R-channel receiver input.

 

Degradation in accuracy is negligible when using an Agilent 8642A/B, 8656B,

10

or E4432B RF signal generator as the LO source.

Page 10
Image 10
Agilent Technologies Agilent 8753ES Specifications, Test-Port Output Characteristics7, Test-Port Input Characteristics