Agilent 8753ET Specifications
Test Port Input Characteristics
Frequency characteristics Range
Option 006
Resolution
Stability4
Option 1D5
Accuracy
Power range
Resolution
Level accuracy1
Level linearity1
Standard
Option 004
Impedance
300 kHz to 3 GHz
3 GHz to 6 GHz
Spectral purity 2nd harmonic2
Option 002 3rd harmonic3
Option 002
Nonharmonic spurious
300 kHz to 3 GHz
300 kHz to 6 GHz
1 Hz
±7.5 ppm (0° to 55° C) ±3 ppm/year
±0.05 ppm (0° to 55° C) ±0.5 ppm/year
±10 ppm at 25° ±5° C
0.01dB ±1.0 dB
±0.5 dB ±0.2 dB ±0.5 dB
±0.2 dB ±0.5 dB
50 Ω (nominal)
>18 dB RL (< 1.28 SWR) >14 dB RL (<1.50 SWR)
Frequency range | 300 kHz to 3 GHz |
Option 006 | 300 kHz to 6 GHz |
Average noise level |
|
300 kHz to 3 GHz | |
300 kHz to 3 GHz | |
300 kHz to 3 GHz | |
| typical |
3 GHz to 6 GHz | |
3 GHz to 6 GHz | |
3 GHz to 6 GHz | |
| typical |
Maximum input level | +10 dBm reflection port |
| 0 dBm transmission port |
Damage level | >+26 dBm or 35 VDC |
Impedance | 50 Ω (nominal) |
Frequency response5 |
|
300 kHz to 3 GHz | ±1.0 dB |
3 GHz to 6 GHz | ±2.0 dB |
Harmonics (Option 002) |
|
2nd harmonic2 | |
| |
| |
3rd harmonic3 | |
| |
| |
Harmonic measurement accuracy5 | |
300 kHz to 3 GHz | ±1.5 dB |
3 GHz to 6 GHz | ±3 dB (Option 006) |
Harmonic measurement dynamic range4 | |
Frequency offset mode6 |
|
Frequency range | 300 kHz to 3 GHz |
Option 006 | 300 kHz to 6 GHz |
| |
Power level |
|
300 kHz to 3 GHz | 0 to |
3 GHz to 6 GHz | 0 to |
Mixer related | |
|
LO Spectral purity |
|
Maximum spurious input | |
Residual FM | <20 kHz |
LO frequency accuracy6 | ±1 MHz at nominal frequency |
1. | Relative to |
| for 8753ET with Option 004. |
2. | 16 MHz to 3 GHz. |
3. | 16 MHz to 2 GHz. |
4. | Typical performance. |
5. | 25° C ± 5° C. |
6. | The Agilent 8753ET source characteristics and measurement accuracy in this |
| mode are dependent on the stability of the external LO source. The RF source |
| tracks the LO to maintain a stable IF signal at the |
14 | Degradation in accuracy is negligible when using an Agilent 8642A/B, 8656B, |
or E4432B RF signal generator as the LO source. |