Agilent Technologies 8753ES manual Agilent 8753ET Specifications

Models: 8753ES 8753ET

1 28
Download 28 pages 56.53 Kb
Page 14
Image 14

Agilent 8753ET Specifications

Test-Port Output Characteristics

Test Port Input Characteristics

Frequency characteristics Range

Option 006

Resolution

Stability4

Option 1D5

Accuracy

Power range

Resolution

Level accuracy1

Level linearity1

Standard

-20 to -15 dBm -15 to 0 dBm 0 to +5 dBm

Option 004

-15 to +5 dBm +5 to +10 dBm

Impedance

300 kHz to 3 GHz

3 GHz to 6 GHz

Spectral purity 2nd harmonic2

Option 002 3rd harmonic3

Option 002

Nonharmonic spurious

300 kHz to 3 GHz

300 kHz to 6 GHz

1 Hz

±7.5 ppm (0° to 55° C) ±3 ppm/year

±0.05 ppm (0° to 55° C) ±0.5 ppm/year

±10 ppm at 25° ±5° C -20 to +5 dBm

-85 to +10 dBm (with Option 004)

0.01dB ±1.0 dB

±0.5 dB ±0.2 dB ±0.5 dB

±0.2 dB ±0.5 dB

50 Ω (nominal)

>18 dB RL (< 1.28 SWR) >14 dB RL (<1.50 SWR)

<-40 dBc at 0 dBm4 <-50 dBc at -10 dBm4

<-25 dBc at maximum output power

<-40 dBc at 0 dBm <-50 dBc at -10 dBm

<-25 dBc at maximum output power

Frequency range

300 kHz to 3 GHz

Option 006

300 kHz to 6 GHz

Average noise level

 

300 kHz to 3 GHz

<-90 dBm (3 kHz BW)

300 kHz to 3 GHz

<-110 dBm (10 Hz BW)

300 kHz to 3 GHz

<-120 dBm (10 Hz BW)

 

typical

3 GHz to 6 GHz

<-85 dBm (3 kHz BW)

3 GHz to 6 GHz

<-105 dBm (10 Hz BW)

3 GHz to 6 GHz

<-114 dBm (10 Hz BW)

 

typical

Maximum input level

+10 dBm reflection port

 

0 dBm transmission port

Damage level

>+26 dBm or 35 VDC

Impedance

50 Ω (nominal)

Frequency response5

 

300 kHz to 3 GHz

±1.0 dB

3 GHz to 6 GHz

±2.0 dB

Harmonics (Option 002)

 

2nd harmonic2

<-15 dBc at +8 dBm

 

<-30 dBc at 0 dBm4

 

<-45 dBc at -15 dBm4

3rd harmonic3

<-30 dBc at +8 dBm

 

<-50 dBc at 0 dBm4

 

<-50 dBc at -15 dBm4

Harmonic measurement accuracy5

300 kHz to 3 GHz

±1.5 dB

3 GHz to 6 GHz

±3 dB (Option 006)

Harmonic measurement dynamic range4

-40 dBc (output = -10 dBm, input = <-15 dBm)

Frequency offset mode6

 

Frequency range

300 kHz to 3 GHz

Option 006

300 kHz to 6 GHz

R-channel input requirements

 

Power level

 

300 kHz to 3 GHz

0 to -35 dBm

3 GHz to 6 GHz

0 to -30 dBm

Mixer related

<-30 dBc at +10 dBm

 

<-55 dBc at -10 dBm

LO Spectral purity

 

Maximum spurious input

<-25 dBc

Residual FM

<20 kHz

LO frequency accuracy6

±1 MHz at nominal frequency

1.

Relative to -5 dBm output power for the Agilent 8753ET; -10 dBm output power

 

for 8753ET with Option 004.

2.

16 MHz to 3 GHz.

3.

16 MHz to 2 GHz.

4.

Typical performance.

5.

25° C ± 5° C.

6.

The Agilent 8753ET source characteristics and measurement accuracy in this

 

mode are dependent on the stability of the external LO source. The RF source

 

tracks the LO to maintain a stable IF signal at the R-channel receiver input.

14

Degradation in accuracy is negligible when using an Agilent 8642A/B, 8656B,

or E4432B RF signal generator as the LO source.