Agilent 8753ES (50-ohm systems)

Type-N test ports

Configuration

 

Network analyzer

8753ES

 

Standard and Option 006

Calibration kit

85032B

Test-port cables

11857D

System dynamic range

 

30 kHz to 50 kHz

70 dB1

50 kHz to 300 kHz

90 dB1

300 kHz to 16 MHz

100 dB

16 MHz to 3 GHz

110 dB

Option 014

108 dB

3 GHz to 6 GHz

105 dB

Option 014

103 dB

Reflection uncertainty

Magnitude

Transmission uncertainty

Magnitude

Measurement Port Characteristics

Frequency Range

Corrected

30 kHz–

300 kHz–

1.3 GHz–

3

GHz–

 

300 kHz2

1.3 GHz

3 GHz

6

GHz

 

 

 

 

 

Directivity

50 dB

50 dB

47 dB

40 dB

 

 

 

 

 

Source match

49 dB

42 dB

36 dB

31 dB

 

 

 

 

 

Load match

50 dB

50 dB

47 dB

40 dB

 

 

 

 

 

Reflection

±(0.005 dB

±(0.009 dB

±(0.019 dB

±(0.070 dB

tracking

+0.02 dB/°C)

+0.01 dB/°C)

+0.02 dB/°C)

+0.03 dB/°C)

 

 

 

 

 

Transmission

±(0.014 dB

±(0.013 dB

±(0.026 dB

±(0.065 dB

tracking

+0.02 dB/°C)

+0.01 dB/°C)

+0.02 dB/°C)

+0.03 dB/°C)

 

 

 

 

 

 

Phase

Phase

1.

Typical below 300 kHz.

5

2.

Typical performance.

Page 5
Image 5
Agilent Technologies 8753ET, 8753ES manual Type-N test ports, Calibration kit 85032B Test-port cables 11857D