External source mode1 (CW time sweep only)
Frequency range | 300 kHz to 6 GHz |
Power level | 0 to |
Spectral purity |
|
Maximum | |
spurious input |
|
Residual FM | <20 kHz |
Typical settling time | 500 ms (automatic) |
| 50 ms (manual) |
Frequency readout accuracy | 0.1% (automatic) |
Input frequency accuracy requirement2 | |
Manual | |
Accuracy | (see magnitude and phase |
| characteristics) |
Reference level |
|
Range | ±500 dB |
Resolution | 0.001 dB |
Stability2 |
|
30 kHz to 3 GHz | 0.02 dB/° C |
3 GHz to 6 GHz | 0.04 dB/° C |
Phase Characteristics
Dynamic accuracy | (10 Hz IF BW) |
Range | ±180° |
Magnitude Characteristics
Dynamic accuracy | (10 Hz IF BW) |
|
| Display resolution | 0.01°/division | |
|
| Marker resolution3 | 0.01° | |
|
| Trace noise |
| |
|
| (+5 dBm at | ||
|
| 300 kHz to 3 GHz | <0.038° rms | |
|
| 3 GHz to 6 GHz | <0.070° rms | |
Display resolution | 0.001 dB/division | Reference level |
| |
Marker resolution3 | 0.001 dB | Range | ||
Trace noise |
| Resolution | 0.01° | |
(+5 dBm at | Stability |
| ||
300 kHz to 3 GHz | <0.006 dB rms | 30 kHz to 3 GHz | 0.05°/° C | |
3 GHz to 6 GHz | <0.010 dB rms | 3 GHz to 6 GHz | 0.20°/° C | |
|
| Polar characteristics |
| |
|
| Range | 10 x | |
1. See the Agilent 8753ES descriptions and options for a functional description. |
| full scale | ||
Reference | ±500 units | |||
Measurement accuracy is dependent on the stability of the input signal. | ||||
|
|
2.Typical performance.
3.Marker resolution for magnitude; phase and delay is dependent upon measured
value. Resolution is limited to five digits. | 11 |