External source mode1 (CW time sweep only)

Frequency range

300 kHz to 6 GHz

R-channel input requirements (typical)

Power level

0 to -25 dBm

Spectral purity

 

Maximum

<-30 dBc

spurious input

 

Residual FM

<20 kHz

Typical settling time

500 ms (automatic)

 

50 ms (manual)

Frequency readout accuracy

0.1% (automatic)

Input frequency accuracy requirement2

Manual

-0.5 to 5 MHz

Accuracy

(see magnitude and phase

 

characteristics)

Reference level

 

Range

±500 dB

Resolution

0.001 dB

Stability2

 

30 kHz to 3 GHz

0.02 dB/° C

3 GHz to 6 GHz

0.04 dB/° C

Phase Characteristics

Dynamic accuracy

(10 Hz IF BW)

Range

±180°

Magnitude Characteristics

Dynamic accuracy

(10 Hz IF BW)

 

 

Display resolution

0.01°/division

 

 

Marker resolution3

0.01°

 

 

Trace noise

 

 

 

(+5 dBm at test-port, ratio measurement, 3 kHz BW)

 

 

300 kHz to 3 GHz

<0.038° rms

 

 

3 GHz to 6 GHz

<0.070° rms

Display resolution

0.001 dB/division

Reference level

 

Marker resolution3

0.001 dB

Range

-180° to +180°

Trace noise

 

Resolution

0.01°

(+5 dBm at test-port, ratio measurement, 3 kHz BW)

Stability

 

300 kHz to 3 GHz

<0.006 dB rms

30 kHz to 3 GHz

0.05°/° C

3 GHz to 6 GHz

<0.010 dB rms

3 GHz to 6 GHz

0.20°/° C

 

 

Polar characteristics

 

 

 

Range

10 x 10–12to 1000 units

1. See the Agilent 8753ES descriptions and options for a functional description.

 

full scale

Reference

±500 units

Measurement accuracy is dependent on the stability of the input signal.

 

 

2.Typical performance.

3.Marker resolution for magnitude; phase and delay is dependent upon measured

value. Resolution is limited to five digits.

11

Page 11
Image 11
Agilent Technologies 8753ET Phase Characteristics, Magnitude Characteristics, External source mode1 CW time sweep only