System Dynamic Range |
|
30 kHz to 50 kHz | 68 dB1 |
50 kHz to 300 kHz | 90 dB1 |
300 kHz to 16 MHz | 96 dB |
16 MHz to 3 GHz | 106 dB |
Reflection uncertainty
Magnitude
Transmission uncertainty
Magnitude
Measurement Port Characteristics
Frequency Range
Corrected | 30 | 300 | 1.3 | |
Directivity | 48 dB | 48 dB |
| 43 dB |
Source match | 47 dB | 41 dB |
| 35 dB |
Load match | 48 dB | 48 dB |
| 43 dB |
Reflection | ±(0.004 dB | ±(0.010 dB | ±(0.019 dB | |
tracking | +0.02 dB/°C) | +0.01 dB/°C) | +0.02 dB/°C) | |
Transmission | ±(0.018 dB | ±(0.016 dB | ±(0.033 dB | |
tracking | +0.02 dB/°C) | +0.01 dB/°C) | +0.02 dB/°C) | |
Uncorrected2 |
|
|
|
|
Directivity | 20 dB* | 35 dB |
| 30 dB |
Source match | 16 dB** | 16 dB |
| 16 dB |
Load match | 15 dB** | 18 dB |
| 16 dB |
Reflection | ±2.5 dB | ±1 dB |
| ±1 dB |
tracking |
|
|
|
|
Transmission | ±2.5 dB | ±1 dB |
| ±1 dB |
tracking |
|
|
|
|
Crosstalk | 90 dB*** | 100 dB |
| 100 dB |
*15 dB, 30 to 50 kHz | **10 dB, 30 to 50 kHz | ***60 dB, 30 to 50 kHz |
Phase
Phase
1. | Typical below 300 kHz. | 7 |
2. | Typical performance. |