Agilent 8753ES (75-Ohm Systems)

Type-N test ports

Configuration

 

Network analyzer

8753ES Option 075

Calibration kit

85036B

Test-port cables

11857B

System Dynamic Range

 

30 kHz to 50 kHz

68 dB1

50 kHz to 300 kHz

90 dB1

300 kHz to 16 MHz

96 dB

16 MHz to 3 GHz

106 dB

Reflection uncertainty

Magnitude

Transmission uncertainty

Magnitude

Measurement Port Characteristics

Frequency Range

Corrected

30 kHz–300 kHz2

300 kHz–1.3 GHz

1.3 GHz–3 GHz

Directivity

48 dB

48 dB

 

43 dB

Source match

47 dB

41 dB

 

35 dB

Load match

48 dB

48 dB

 

43 dB

Reflection

±(0.004 dB

±(0.010 dB

±(0.019 dB

tracking

+0.02 dB/°C)

+0.01 dB/°C)

+0.02 dB/°C)

Transmission

±(0.018 dB

±(0.016 dB

±(0.033 dB

tracking

+0.02 dB/°C)

+0.01 dB/°C)

+0.02 dB/°C)

Uncorrected2

 

 

 

 

Directivity

20 dB*

35 dB

 

30 dB

Source match

16 dB**

16 dB

 

16 dB

Load match

15 dB**

18 dB

 

16 dB

Reflection

±2.5 dB

±1 dB

 

±1 dB

tracking

 

 

 

 

Transmission

±2.5 dB

±1 dB

 

±1 dB

tracking

 

 

 

 

Crosstalk

90 dB***

100 dB

 

100 dB

*15 dB, 30 to 50 kHz

**10 dB, 30 to 50 kHz

***60 dB, 30 to 50 kHz

Phase

Phase

1.

Typical below 300 kHz.

7

2.

Typical performance.

Page 7
Image 7
Agilent Technologies 8753ET manual Agilent 8753ES 75-Ohm Systems, Network analyzer