Agilent Technologies 8753ET, 8753ES manual Agilent 85047A S-Parameter Test Set, Specifications

Models: 8753ES 8753ET

1 28
Download 28 pages 56.53 Kb
Page 23
Image 23

Agilent 85047A S-Parameter Test Set

The 85047A S-parameter test set provides the capability to simultaneously measure the reflec- tion and transmission characteristics of two-port devices in either direction with a single connec- tion. This test set includes a frequency doubler that can be switched in by an 8753B/C Option 006 to measure 3 MHz to 6 GHz in a single sweep or switched out to measure 300 kHz to 3 GHz in

a single sweep. The 8753ES Option 011 does not use the frequency doubler, so the full 300 kHz to 6 GHz range is available. This test set exhibits

<5 dB insertion loss between the RF input and the test ports for as high as 15 dBm at the test port, and also includes a programmable step attenuator. There are two rear panel BNC outputs. One pro- vides a TTL signal which indicates the result of

a limit test. The second TTL output is controlled from the 8753ES test sequence function.

Specifications

Impedance

50 ohms

Frequency range

300 kHz to 3 GHz and

 

3 GHz to 6 GHz

 

(8753B/C);

 

300 kHz to 6 GHz

 

(8753D/E/ES Opt. 006)

Directivity1

 

300 kHz to 1.3 GHz

35 dB2

1.3 GHz to 3 GHz

30 dB

3 GHz to 6 GHz

25 dB

Typical tracking1

 

Transmission magnitude, phase3

300 kHz to 3 GHz

±1.5 dB, ±10°

3 GHz to 6 GHz

+0.5, -2.5 dB, ±20°

Reflection magnitude, phase3

 

300 kHz to 3 GHz

±1.5 dB, ±10°

3 GHz to 6 GHz

±1.5 dB, ±20°

Source match1

 

300 kHz to 1.3 GHz

20 dB

1.3 GHz to 3 GHz

16 dB

3 GHz to 6 GHz

14 dB

Normal insertion loss

 

Input to port 1,2

 

300 kHz to 3 GHz

4.0 dB +0.8 dB/GHz

3 GHz to 6 GHz

17.5 dB +0.8 dB/GHz

Input to R

 

300 kHz to 3 GHz

19 dB +0.5 dB/GHz

3 GHz to 6 GHz

34 dB +0.5 dB/GHz

Port 1,2 to A,B

16 dB

Typical isolation4

 

300 kHz to 3 GHz

100 dB

3 GHz to 6 GHz

90 dB

Test port switch

±0.03 dB; ±0.01 dB4

repeatability

 

Maximum operating level

+20 dBm

Damage level

+30 dBm

RF attenuator range

70 dB (10 dB steps)

DC bias range

±30 VDC, 200 mA,

 

no degradation in RF specs,

 

1A max.

RF connectors

 

Port 1,2

7 mm precision

All others

50 ohm type-N(f)

Dimensions

90 mm H x 432 mm W x

 

553 mm D

Weight

10 kg (22 lb)

A standard 85047A test set contains a solid-state transfer switch, which allows continuous switching of power from port 1 to port 2 for full two-port error correction.

Agilent 85047A Schematic

1.This can be greatly improved with accuracy enhancement.

2.Some degradation at environmental extremes below 600 kHz.

3.Degrees, specified as deviation from linear phase.

4. Typical performance.

23

Page 23
Image 23
Agilent Technologies 8753ET, 8753ES manual Agilent 85047A S-Parameter Test Set, Specifications