Agilent 8753ET (50-ohm systems)

Type-N test ports

Configuration

 

 

 

Network analyzer

8753ET Standard,

 

 

Option 006, or Option 004

Calibration kit

 

85032B

 

Test port cable

 

Part number 8120-4781

System dynamic range

 

 

300 kHz to 16 MHz

100 dB

 

16 MHz to 3 GHz

110 dB

 

3 GHz to 6 GHz

 

105 dB

 

Measurement Port Characteristics

 

Frequency Range

 

 

 

 

 

 

 

Uncorrected

300 kHz–1.3 GHz

1.3 GHz–3.0 GHz

3.0 GHz–6.0 GHz

 

 

 

 

Directivity

30 dB

24 dB

19 dB

 

 

 

 

Source match4

25 dB

20 dB

14 dB

 

 

 

 

Load match

24 dB

19 dB

16 dB

 

 

 

 

Reflection

±1.0 dB

±1.0 dB

±2.0 dB

tracking

 

 

 

 

 

 

 

Transmission

±1.5 dB

±1.5 dB

±2.5 dB

tracking

 

 

 

 

 

 

 

Crosstalk

100 dB

100 dB

90 dB

 

 

 

 

Measurement Port Characteristics (continued)

Corrected

Reflection

 

 

 

measurements1

300 kHz–1.3 GHz

1.3 GHz–3.0 GHz

3.0 GHz–6.0 GHz

 

 

 

 

Directivity

50 dB

47 dB

40 dB

 

 

 

 

Source match

42 dB

36 dB

31 dB

 

 

 

 

Load match

 

 

 

One-port cal

24 dB

19 dB

16 dB

Enhanced-

24 dB

19 dB

16 dB

reflection cal

 

 

 

 

 

 

 

Reflection

±(0.009 dB

±(0.019 dB

±(0.07 dB

tracking

+0.01 dB/°C)

+0.02 dB/°C)

+0.03 dB/°C)

 

 

 

 

Transmission

 

 

 

measurements2

 

 

 

 

 

 

 

Source match

 

 

 

Enhanced-

42 dB

36 dB

31 dB

response cal

 

 

 

Response-only cal3 25 dB

20 dB

14 dB

 

 

 

Transmission tracking

 

 

Enhanced-

±(0.006 dB

±(0.018 dB

±(0.054 dB

response cal

+0.01 dB/°C)

+0.02 dB/°C)

+0.03 dB/°C)

Response-only cal

±(0.033 dB

±(0.1 dB

±(0.27 dB

 

+0.01 dB/°C)

+0.02 dB/°C)

+0.03 dB/°C)

 

 

 

 

Transmission uncertainty (enhanced-response calibration)

Transmission uncertainty (response calibration)

Reflection uncertainty (one-port calibration)

1.One-port or enhanced-response calibration.

2.Enhanced-response or response-only calibration.

3.2 dB less with Option 004.

4.Option 004 may degrade uncorrected source match as much as 2 dB.

9

Page 9
Image 9
Agilent Technologies 8753ES manual Agilent 8753ET 50-ohm systems, Part number, KHz to 16 MHz 100 dB MHz to 3 GHz 110 dB