Agilent 8753ES Option 011 Specifications
Test Port Output Characteristics
Frequency characteristics |
|
Range | 300 kHz to 3 GHz |
Option 006 | 30 kHz to 6 GHz |
Resolution | 1 Hz |
Stability | typically ±7.5 ppm 0° to 55° C |
| typically ±3 ppm/year |
Option 1D5 | typically ±0.05 ppm 0° to 55° C |
| typically ±0.5 ppm/year |
Accuracy | ±10 ppm at 25° ± 5° C |
Power range | |
Option 006 | |
Resolution | 0.01 dB |
Level accuracy1,2 | ±1.0 dB |
Level linearity1,2,5 | ±0.25 dB, |
| ±0.5 dB, +15 to +20 dBm |
Impedance | 50 Ω nominal |
300 kHz to 3 GHz2 >16 dB RL (<1.38 SWR) | |
300 kHz to 6 GHz | >14 dB RL (<1.50 SWR) |
Spectral purity |
|
2nd harmonic3 | |
| |
Option 002 | |
3rd harmonic4 | |
| |
Option 002 | |
Nonharmonic spurious6 |
|
Mixer related | |
|
Test Port Input Characteristics Option 011
Frequency range | 300 kHz to 3 GHz |
Option 006 | 30 kHz to 6 GHz |
Average noise level2 |
|
50 kHz to 3 GHz | |
50 kHz to 3 GHz | |
50 kHz to 3 GHz | |
3 to 6 GHz | |
3 to 6 GHz | |
3 to 6 GHz | |
Maximum input level | 0 dBm |
Damage level | 20 dBm or 25 VDC |
Impedance | 50 ohms nominal |
300 kHz to 2 MHz | ≥20 dB RL |
2 MHz to 1.3 GHz | ≥24 dB RL |
1.3 GHz to 3 GHz | ≥19 dB RL |
Option 006 | ≥15 dB RL6 |
3 GHz to 6 GHz | |
Frequency response |
|
300 kHz to 3 GHz | ±1.0 dB |
3 GHz to 6 GHz | ±2.0 dB |
Harmonics (Option 002) |
|
2nd harmonic3 | |
| |
| |
3rd harmonic4 | |
| |
| |
Harmonic measurement accuracy | |
16 MHz to 3 GHz | ±1.5 dB |
3 GHz to 6 GHz Option 006 | ±3 dB |
Harmonic measurement dynamic range
-40 dBc6 (output = -10 dBm, input <-15 dBm)
| 1. | +10 dBm output power for Agilent 8753ES Option 011. |
| 2. | Typical below 300 kHz. |
| 3. | 16 MHz to 3 GHz. |
| 4. | 16 MHz to 2 GHz. |
| 5. | For Agilent 8753ES Option 011 and Option 006, linearity is specified for the |
12 |
| ranges of |
6. | Typical performance. |