276
S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb
Chapter 5, Advanced Operations
Status Reporting
Calibration Status Register Group

The Calibration Status Register Group contains information about the Test Set’s

hardware. This status group is accessed using the STATus commands. The

Calibration Status Register Group uses 16-bit registers and includes a Condition

Register, Transition Filters, an Event Register , an Enable Register, and a Summary

Message. Refer to the “Status Reporting Structure Overview” on page 239 for a

discussion of status register operation. Figure 14 shows the structure and STATus

commands for the Calibration Status Register Group.

Figure 14 Calibration Status Register Group

0
:STATus: CALibration: ENABle?
<integer>
:STATus: CALibration: EVENt ?
:STATus: CALibration
:PTRansition
:NTRansition
?
<integer>
?
<integer>
:STATus: CALibration: CONDition ?
Summary Message Bit
Condition
Register Transition
Filter Registers Event
Register Enable
Register
1
2
3
4
5
6
7
8
9
10
11
12
13
14
Logical OR
15
ch4drw11.drw
(to bit 8 of Questionable
Data/Signal Register Group
Condition Register