Chapter 5, Advanced Operations
Instrument Initialization
The *RST Reset condition in the Test Set was specifically designed to configure the instruments for manual testing of an FM radio. The *RST Reset default display screen is the RX TEST screen. Other operational characteristics are also affected by the *RST reset as follows:
All pending operations are aborted.
Measurement triggering is set to TRIG:MODE:SETT FULL;RETR REP.
Any previously received Operation Complete command (*OPC) is cleared.
Any previously received Operation Complete query command (*OPC?) is cleared.
The Test Set’s display screen is in the UNLOCKED state.
The Power-up self-test diagnostics are not performed.
The Contents of the SAVE/RECALL registers are not affected.
Calibration data is not affected.
The GPIB interface is not reset (any pending Service Request is not cleared).
All Enable registers are unaffected: Service Request, Standard Event, Communicate,
Hardware #1, Hardware #2, Operation, Calibration,Call Processing and Questionable
Data/Signal.
All Negative Transition Filter registers are unaffected: Communicate, Hardware #1,
Hardware #2, Operation, Calibration,Call Processing and Questionable Data/Signal.
All Positive Transition Filter registers are unaffected: Communicate, Hardware #1,
Hardware #2, Operation, Calibration,Call Processing and Questionable Data/Signal.
The contents of the RAM memory are unaffected.
The contents of the Output Queue are unaffected.
The contents of the Error Queue are unaffected.