557
Error 202 Settings lost due to rtl
Indicates that a setting associated with a hard local control (see IEEE 488.2, 5.6.1.5) was
lost when the device changed to LOCS from REMS or to LWLS from RWLS.
Error 210 Trigger error
Error 211 Trigger ignored
Indicates that a GET, *TRG, or triggering signal was received and recognized by the
device but was ignored because of device timing considerations. For example, the device
was not ready to respond.
Error 212 Arm ignored
Indicates that an arming signal was received and recognized by the device but was
ignored.
Error 213 Init ignored
Indicates that a request for a measurement initiation was ignored as another measurement
was already in progress.
Error 214 Trigger deadlock
Indicates that the trigger source for the initiation of a measurement is set to GET and
subsequent measurement query is received. The measurement cannot be started until a
GET is received, but the GET would cause an INTERRUPTED error.
Error 215 Arm deadlock
Indicates that the arm source for the initiation of a measurement is set to GET and
subsequent measurement query is received. The measurement cannot be started until a
GET is received, but the GET would cause an INTERRUPTED error.
Error 220 Parameter error
Indicates that a program data element related error occurred.