Programming the Status Register System

Status Groups

Status Groups

The Standard Operation Status Group and the Data Questionable Status Group consist of the registers listed below. The Standard Event Status Group is similar but does not have negative or positive transition filters or a condition register.

Condition

Register A condition register continuously monitors the hardware and firmware status of the signal generator. There is no latching or buffering for a condition register; it is updated in real time.

Negative

Transition

FilterA negative transition filter specifies the bits in the condition register that will set corresponding bits in the event register when the condition bit changes from 1 to 0.

Positive Transition

FilterA positive transition filter specifies the bits in the condition register that will set corresponding bits in the event register when the condition bit changes from 0 to 1.

Event

Register An event register latches transition events from the condition register as specified by the positive and negative transition filters. Once the bits in the event register are set, they remain set until cleared by either querying the register contents or sending the *CLS command.

Event

Enable

Register An enable register specifies the bits in the event register that generate the summary bit. The signal generator logically ANDs corresponding bits in the event and enable registers and ORs all the resulting bits to produce a summary bit. Summary bits are, in turn, used by the Status Byte Register.

A status group is a set of related registers whose contents are programmed to produce status summary bits. In each status group, corresponding bits in the condition register are filtered by the negative and positive transition filters and stored in the event register. The contents of the event register are logically ANDed with the contents of the enable register and the result is logically ORed to produce a status summary bit in the Status Byte Register.

156

Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide

Page 166
Image 166
Agilent Technologies N5183A MXG, N5181A/82A, E8663B manual Status Groups

N5183A, N5183A MXG, E8663B, N5181A/82A specifications

Agilent Technologies is renowned for its innovative solutions in electronic test and measurement equipment. Among its offerings are several signal generators including the N5181A, N5182A, E8663B, and N5183A MXG. These models are distinguished not only by their performance but also by their versatility across various applications in communications, aerospace, and electronics.

The Agilent N5181A and N5182A, part of the MXG family, are highly versatile signal generators known for their exceptional frequency performance and flexibility. The N5181A operates from 100 kHz to 6 GHz, while the N5182A extends that range up to 12 GHz. They provide high fidelity signals with low phase noise, making them ideal for the development and testing of RF components and systems. These generators support a wide variety of modulation formats, including AM, FM, PM, and pulse modulation, catering to diverse application needs.

The E8663B, meanwhile, is designed for the range of 250 kHz to 3 GHz and is also recognized for its high-performance features. Its built-in capabilities for modulation make it effective for testing wireless devices, ensuring that signals can be simulated accurately in both laboratory and field environments. It is especially beneficial for users requiring a straightforward and efficient solution with high reliability.

The N5183A MXG signal generator enhances the lineup with frequency coverage up to 6 GHz and advanced capabilities. It integrates various modulation capabilities while ensuring high signal integrity. Its architecture is tailored for both production test environments and research applications, providing users with the flexibility to adapt to changing testing requirements.

Common characteristics across these models include a user-friendly interface that simplifies configuration and operation. They are often equipped with LAN and USB interfaces for easy remote control and integration into automated test systems. The robustness of these generators allows them to perform reliably in challenging environments, making them essential tools in laboratories, manufacturing floors, and field testing scenarios.

In summary, Agilent's signal generators, including the N5181A, N5182A, E8663B, and N5183A MXG, represent a blend of advanced technology, flexibility, and precision. These instruments are vital in facilitating the evolution of cutting-edge communication technologies, ensuring that designers and engineers can confidently meet the demands of modern electronics.