Using IO Interfaces

Using RS-232 (ESG, PSG, and E8663B Only)

Using RS-232 (ESG, PSG, and E8663B Only)

NOTE The RS- 232 serial interface is available on the ESG signal generators.

The PSG and E8663B’s AUXILIARY INTERFACE connector is compatible with ANSI/EIA232 (RS- 232) serial connection but GPIB and LAN are recommended for making faster measurements and when downloading files. Refer to the User’s Guide.

The RS- 232 serial interface can be used to communicate with the signal generator. The RS- 232 connection was once standard on most PCs but has now been replaced by USB. RS- 232 can be connected to the signal generator’s rear- panel connector using the cable described in Table 2- 2 on page 47. Many functions provided by GPIB, with the exception of indefinite blocks, parallel polling, serial polling, GET, non- SCPI remote languages, SRQ, and remote mode are available using the RS- 232 interface.

The serial port sends and receives data one bit at a time, therefore RS- 232 communication is slow. The data transmitted and received is usually in ASCII format with SCPI commands being sent to the signal generator and ASCII data returned.

The following sections contain information on selecting and connecting IO libraries and RS- 232 interface hardware on the signal generator to a computer’s RS- 232 connector.

“Selecting IO Libraries for RS- 232” on page 45

“Setting Up the RS- 232 Interface” on page 46

“Verifying RS- 232 Functionality” on page 48

Selecting IO Libraries for RS-232

The IO libraries can be downloaded from the National Instrument website, http://www.ni.com, or Agilent’s website, http://www.agilent.com. The following is a discussion on these libraries.

CAUTION

Because of the potential for portability problems, running Agilent SICL without the

 

VISA overlay is not recommended by Agilent Technologies.

 

 

HP Basic

The HP Basic language has an extensive IO library that can be used to control the

 

signal generator over the RS- 232 interface. This library has many low level

 

functions that can be used in BASIC applications to control the signal generator

 

over the RS- 232 interface.

VISA

VISA is an IO library used to develop IO applications and instrument drivers that

 

comply with industry standards. It is recommended that the VISA library be used

 

for programming the signal generator. The NI- VISA and Agilent VISA libraries are

 

similar implementations of VISA and have the same commands, syntax, and

 

functions. The differences are in the lower level IO libraries used to communicate

 

over the RS- 232; NI- 488.2 and SICL respectively.

NI- 488.2

NI- 488.2 IO libraries can be used to develop applications for the RS- 232 interface.

 

See National Instrument’s website for information on NI- 488.2.

Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide

45

Page 55
Image 55
Agilent Technologies N5183A, N5181A/82A manual Using RS-232 ESG, PSG, and E8663B Only, Selecting IO Libraries for RS-232

N5183A, N5183A MXG, E8663B, N5181A/82A specifications

Agilent Technologies is renowned for its innovative solutions in electronic test and measurement equipment. Among its offerings are several signal generators including the N5181A, N5182A, E8663B, and N5183A MXG. These models are distinguished not only by their performance but also by their versatility across various applications in communications, aerospace, and electronics.

The Agilent N5181A and N5182A, part of the MXG family, are highly versatile signal generators known for their exceptional frequency performance and flexibility. The N5181A operates from 100 kHz to 6 GHz, while the N5182A extends that range up to 12 GHz. They provide high fidelity signals with low phase noise, making them ideal for the development and testing of RF components and systems. These generators support a wide variety of modulation formats, including AM, FM, PM, and pulse modulation, catering to diverse application needs.

The E8663B, meanwhile, is designed for the range of 250 kHz to 3 GHz and is also recognized for its high-performance features. Its built-in capabilities for modulation make it effective for testing wireless devices, ensuring that signals can be simulated accurately in both laboratory and field environments. It is especially beneficial for users requiring a straightforward and efficient solution with high reliability.

The N5183A MXG signal generator enhances the lineup with frequency coverage up to 6 GHz and advanced capabilities. It integrates various modulation capabilities while ensuring high signal integrity. Its architecture is tailored for both production test environments and research applications, providing users with the flexibility to adapt to changing testing requirements.

Common characteristics across these models include a user-friendly interface that simplifies configuration and operation. They are often equipped with LAN and USB interfaces for easy remote control and integration into automated test systems. The robustness of these generators allows them to perform reliably in challenging environments, making them essential tools in laboratories, manufacturing floors, and field testing scenarios.

In summary, Agilent's signal generators, including the N5181A, N5182A, E8663B, and N5183A MXG, represent a blend of advanced technology, flexibility, and precision. These instruments are vital in facilitating the evolution of cutting-edge communication technologies, ensuring that designers and engineers can confidently meet the demands of modern electronics.